29
The warning will flash for a few seconds and the unit will turn itself
off. Replace the battery by sliding off the battery cover at the back
of the unit, unclip the exhausted battery and replace with a new
battery of the correct type.
Note: if the battery is not to be used for long periods, the battery
should be removed to prevent damage to the instrument by possible
leakage.
Calibration Check
The MicroDL is calibrated to read in litres at body temperature,
barometric pressure, and saturated with water vapour (BTPS) for
FEV1 and FVC. The units of PEF are litres per second BTPS or litres
per minute BTPS depending upon the configuration in the PC
software.
The calibration should remain stable indefinitely, unless the
transducer is physically damaged, and the unit should not require
re-calibration.
However, as a check on the correct functioning of the unit we do
recommend that the calibration is checked periodically.
To check the calibration of the unit a large syringe should be used.
This should have a 30mm outlet and be of at least three litres in
capacity and preferably six litres.
A 3 Litre Syringe is available from CareFusion as Cat. No. 36-
SM2125. The MicroDL should be connected to this syringe with the
Summary of Contents for MicroDL
Page 1: ...1 MicroDL Operating Manual English Bedienungsanleitung Deutsch...
Page 2: ......
Page 4: ...2 IV 103 104 MicroDL 106 MicroDL SPCS 129 131 131 132 134...
Page 36: ......
Page 40: ...37 MicroDL 1 Fig 1 of enter 3 4 on 2...
Page 73: ...70 MicroDL 1 Fig 1 of enter 3 4 on 2...
Page 106: ......
Page 107: ...103 MicroDL IV MicroDL CareFusion MicroDL Spirometry PC Software SPCS...
Page 108: ...104 SPCS 200 CareFusion B T P S MicroDL 1 MicroDL 2 CareFusion 3 3 4 SPCS...
Page 109: ...105 MicroDL 1 Fig 1 of enter 3 4 on 2...
Page 110: ...106 MicroDL MicroDL SPCS FEV1 FVC PEF FEV1 FVC SPCS MicroDL ON OFF...
Page 111: ...107 SPCS ENTER 1 ENTER...
Page 112: ...108 MicroDL ENTER ENTER...
Page 113: ...109 SPCS ENTER...
Page 114: ...110 1 1999 ENTER ENTER...
Page 115: ...111 ENTER ENTER ENTER MicroDL...
Page 116: ...112 PEF FEV1 FVC SPCS...
Page 117: ...113 PEF PEF SPCS SPCS PEF FEV1 FVC PEF FEV1 FVC...
Page 119: ...115 ENTER EO E1 E4 ENTER ENTER...
Page 120: ...116 2 ENTER 1 1999 ENTER...
Page 121: ...117 ENTER...
Page 122: ...118 ENTER...
Page 123: ...119 PEF FEV1 FVC FEV1 FVC FEV1FVC SPCS ENTER OFF SPCS FEV1 FVC PEF FEV1 FVC OFF...
Page 126: ...122 1 MicroDL...
Page 128: ...124 FEV1 FVC PEF 2 ENTER 2 MicroDL 2 1 1 ENTER...
Page 131: ...127 3 MicroDL 3 65 65 65 ENTER 2 OFF...
Page 132: ...128 4 MicroDL 4 4 ENTER 1 5 MicroDL 5 5...
Page 135: ...131 MicroDL 10 9 PP3 24 CareFusion B T P S...
Page 136: ...132 MicroDL BTPS 30 3 6 36 SM2125 MicroDL FVC...
Page 137: ...133 Sidex 15...
Page 142: ......
Page 143: ......