27
Configuring the MicroDL with SPCS
The MicroDL can be set up using SPCS. First, ensure that the
MicroDL is connected to the PC using the cable supplied. Connect
the nine way ‘D’ connector to the serial port of the PC and the 3 pole
jack plug to the socket in the right hand side of the MicroDL.
To communicate with the PC the MicroDL must be turned on in PC
mode. Depress the ‘enter’ key, turn the unit on, and release the
‘enter’ key when the display shows:
Start SPCS on the PC. Ensure that SPCS is set up to use the COM
port on the computer that the MicroDL is connected to (refer to
online SPCS help). A patient record should then be set up
containing the details to be downloaded to the MicroDL (or an
existing patient selected). These details can then be used to
initialise the unit. Please refer to the online help included with SPCS
for the exact details for performing these operations.
NOTE: Keep the PC out of reach of the patient when connected to
the MicroDL. The MicroDL should only be connected to a computer
that is manufactured in accordance with EN60950 1992/1993 ‘Safety
of Information Technology Equipment including Electrical Business
Equipment’.
Uploading Results
To upload the results stored in the MicroDL, connect the unit to the
PC as described above and start SPCS. When the initial screen is
displayed, click on the ‘upload‘button. The data contained in the
Summary of Contents for MicroDL
Page 1: ...1 MicroDL Operating Manual English Bedienungsanleitung Deutsch...
Page 2: ......
Page 4: ...2 IV 103 104 MicroDL 106 MicroDL SPCS 129 131 131 132 134...
Page 36: ......
Page 40: ...37 MicroDL 1 Fig 1 of enter 3 4 on 2...
Page 73: ...70 MicroDL 1 Fig 1 of enter 3 4 on 2...
Page 106: ......
Page 107: ...103 MicroDL IV MicroDL CareFusion MicroDL Spirometry PC Software SPCS...
Page 108: ...104 SPCS 200 CareFusion B T P S MicroDL 1 MicroDL 2 CareFusion 3 3 4 SPCS...
Page 109: ...105 MicroDL 1 Fig 1 of enter 3 4 on 2...
Page 110: ...106 MicroDL MicroDL SPCS FEV1 FVC PEF FEV1 FVC SPCS MicroDL ON OFF...
Page 111: ...107 SPCS ENTER 1 ENTER...
Page 112: ...108 MicroDL ENTER ENTER...
Page 113: ...109 SPCS ENTER...
Page 114: ...110 1 1999 ENTER ENTER...
Page 115: ...111 ENTER ENTER ENTER MicroDL...
Page 116: ...112 PEF FEV1 FVC SPCS...
Page 117: ...113 PEF PEF SPCS SPCS PEF FEV1 FVC PEF FEV1 FVC...
Page 119: ...115 ENTER EO E1 E4 ENTER ENTER...
Page 120: ...116 2 ENTER 1 1999 ENTER...
Page 121: ...117 ENTER...
Page 122: ...118 ENTER...
Page 123: ...119 PEF FEV1 FVC FEV1 FVC FEV1FVC SPCS ENTER OFF SPCS FEV1 FVC PEF FEV1 FVC OFF...
Page 126: ...122 1 MicroDL...
Page 128: ...124 FEV1 FVC PEF 2 ENTER 2 MicroDL 2 1 1 ENTER...
Page 131: ...127 3 MicroDL 3 65 65 65 ENTER 2 OFF...
Page 132: ...128 4 MicroDL 4 4 ENTER 1 5 MicroDL 5 5...
Page 135: ...131 MicroDL 10 9 PP3 24 CareFusion B T P S...
Page 136: ...132 MicroDL BTPS 30 3 6 36 SM2125 MicroDL FVC...
Page 137: ...133 Sidex 15...
Page 142: ......
Page 143: ......