14 Technical characteristics
A.c. mains power supply
Device with Class I power supply
230 V ± 10% 50/60 Hz
115 V ± 10% 50/60 Hz
Maximum current absorbed
80 mA at 230 V ~ ±10%
Mains protection
Fuse: T 0.5 A
Internal power source
12 V – 1800 mah NiMH
Internal power supply protection
Pico fuse SHF SLO-BLO T 5 A Littelfuse
Applied part
CF type
Defibrillation protection
Internal
Input dynamic
± 300 mV @ 0 Hz. ± 10 mV in the bandwidth
Input impedance
>
100 M
Ω
on each electrode
Common mode rejection
>
100 dB with electrode impedance balanced
Frequency response
0,5
÷
150 Hz (-3dB)
Time constant
3.3 s
Acquisition
11 bit
1000 samples/s/channel printing and filters
500 samples/s/channel in calculation and filters
Resolution 5
µ
V/bit
Leads
&*9: 9$
)9 $0$9
8 acquired, 4 derived (III, aVR, aVL, aVF)
Signal memory
10 seconds for each lead in auto mode
Recording sensitivity
5 – 10 – 20 mm/mV ± 5%, auto*
*2,5 - 5 – 10 – 20 mm/mV ± 5%, dependent on number of channels
printed
Writing system
Thermal printer, 8 dot/mm
Usable print height 108 mm
Print channels
Auto mode 3, 6, 3+ R
Manual mode 3 – 6
Paper feed rate
5 mm/s ± 10%
25 – 50 mm/s ± 5%
Video speed
12,5- 25 – 50 mm/s
Heat sensitive paper
roll: in 120 mm , gridded, length 20 m
Z-fold pack
: in 120x100 mm (width x length) sheets, gridded, length 30
m
Pacemaker recognition
Records impulse in compliance with
IEC 60601-2-51/Ed. standards 1
Filters
Mains interference:
Modified notch digital filter 50 – 60 Hz with response @ 32 Hz (-3dB)