Cardoline 80509531 Technical & Service Manual Download Page 20

20 

AR1200VIEW

 

CARDIOLINE

®

 is an et medical devices SpA brand 

w w w . c a r d i o l i n e . b i z

 

  <<< ar1200 view MEMORY TEST  >>> 

 

  Boot 

 

: CRC OK 

 

  Application 

: CRC OK 

 

  Language 

: CRC OK 

 

  Ram Memory 

: OK 

Info:  The  following  items  of  information  are  printed:  model  identification,  serial  number  of  the  device,  details  of  software 

installed, version and language code, reference of any options installed. 

 
SERVICE 
To enter the self-testing menu, switch on the device, press the MENU key and select and confirm the TOOLS, 

SELF-TEST, SERVICE(2) submenu using the arrow key (DOWN). When the access code is requested, press the 

following keys:       FILTER – AMPLITUDES – SPEED – START and confirm. 
the following tests are available in the service section: 

9.1.2  Mark calibration 
Confirm the submenu MARK CALIBRATION (1), position the mark at a distance from the sensor (visible near the 

exit from the device), and confirm “enter”  
The microprocessor automatically reads the sensor output voltages, voltage with white paper and black mark, and 

then sets the digital trimmer that regulates the photodiode current to the optimal settings and then stores them. 
The mark calibration voltages are shown on the display, as is the numerical position of the digital potentiometer.  

9.1.3  Speed calibration 

Testing and setting 

Select and firm the SPEED CAL. (2) submenus, and the speed to calibrate 
The following messages then appear on the display:  Output PWM0 Output – Freq = xxxx Hz with the possibility 

of selecting from:  test – change – exit. 
Confirming the “test” command enables the paper feed and the print unit emits a 1 impulse per second signal. If 

the speed has to be calibrated, select “change” and increase or decrease the frequency to set the paper feed 

speed entered. When the speed has been set correctly, select and confirm with “exit”. 
 
 
The permitted tolerance values for the various paper feed speeds are as follows: 

  50 mm/s ± -5% 

  25 mm/s ± -5% 

  5 mm/s ± -10% 

9.1.4  Metres of paper printed 
Displays the number of metres of paper printed. 

9.1.5  Firmware upgrades 
Enables upgrading of the instrument’s internal firmware. 
The firmware upgrade menu can be accessed from the main menu, pressing the Shift + S keys together. 

Note: Refer to the “Loader” programme manual for further information. 

9.1.6  Print Screen 
Enable print screen ( for internal use only) 
It is highly recommended to leave this function disabled. 

9.2  Print head alignment 

In case of print head adjustment remove mother board as per § 8.3, then follow instructions provided in table T4, 

repeat adjustment until the best result is obtained. 
 

Summary of Contents for 80509531

Page 1: ...w w c a r d i o l i n e b i z T TE EC CH HN NI IC CA AL L S SE ER RV VI IC CE E M MA AN NU UA AL L E EL LE EC CT TR RO OC CA AR RD DI IO OG GR RA AP PH H M MO OD DE EL L C Ca ar rd di io ol li in ne...

Page 2: ...2 AR1200VIEW CARDIOLINE is an et medical devices SpA brand w w w c a r d i o l i n e b i z Index 0 1 2 3 4 2 3 0 1 4 5 1 6...

Page 3: ...3 AR1200VIEW CARDIOLINE is an et medical devices SpA brand w w w c a r d i o l i n e b i z 2 1 7 2 2 2 3 2 2 2 2 2 4 8 9 0 9 0 9 0 3...

Page 4: ...o store up to 40 full ECG exams with no need to print out immediately on paper paper saving mode ECG measurements option automatic ECG parameter measurement program ECG signal interpretive option a us...

Page 5: ...that the signal is restored use only original electrodes or electrodes responding to IEC and AAMI standards If an electrosurgical scalpel is in use the patient cable should be disconnected from the d...

Page 6: ...MH 12V 2 3 Display Display assembly is composed of frame monochromatic LCD display 320 x 240 pixel 4 7 inch as well as a step up converter 2 4 Keyboard The keyboard card consists of the device functio...

Page 7: ...to the patient input port 8 4 1 2 3 6 7 5 15 9 10 11 12 13 14 Port from connection side Pin 1 IN C2 electrode C2 Pin 2 IN C3 electrode C3 Pin 3 IN C4 electrode C4 Pin 4 IN C5 electrode C5 Pin 5 IN C6...

Page 8: ...nstrument s manual recalling that The leakage current to the casing is measuring between the mains supply circuits and a metal sheet no greater than 20 x 10 cm pressed against the casing of the device...

Page 9: ...nnected to the device to the positive pin of instrument 5 1 a connect all the other cable terminals to the negative pin of instrument 5 1 a record the signal for a few seconds check that the amplitude...

Page 10: ...the signal on leads V1 V6 measure the length of the wave cycle recorded on the paper The results should be as follows Period 50 mm 5 for a feed rate of 50 mm s Period 25 mm 5 for a feed rate of 25 mm...

Page 11: ...3 2 0 0 03 4 2 03 4 3 2 2 2C 3 2 22 2C 7 2C 28 3 22 2 2C 7 2C Note The 0 5 Hz pitch linear phase anti drift filter is always on and cannot be switched off The 50 or 60 Hz filter eliminates modified no...

Page 12: ...illustrated in the picture below Changing mains voltage does not require mains fuses rate modification 5 Warning Changing mains voltage imply the replacement of the mentioned label Only original ones...

Page 13: ...nal polarisation circuit Hybrid circuit for amplification filter and clamp 6 1 6 Motor control This consists of the following parts Stroboscopic sensor control Phase comparator between reference frequ...

Page 14: ...d 6 4 Keyboard card The keyboard consists of 18 function keys 10 alphanumeric keys 1 LED for mains line indication CPU connection through bi directional synchronous serial port Testing the keyboard ca...

Page 15: ...15 AR1200VIEW CARDIOLINE is an et medical devices SpA brand w w w c a r d i o l i n e b i z...

Page 16: ...mains fuses if OK replace motherboard Check efficiency of battery and replace if necessary Check T5A fuse on motherboard Keyboard keys not working The key pressed not enabled for the specific function...

Page 17: ...nd display assembly Verify on display connector J49 presence of LCD data signal level from 0 trough 5 V if present replace display assembly otherwise replace mother board 8 How to dismantle and reasse...

Page 18: ...sembly proceed in inverse order 8 8 Removing the printer assembly table T2b Proceed as follows proceed as in chapter 8 3 for opening the equipment and removing the mother board from the equipment remo...

Page 19: ...pull it off the casing If any adhesive is left on the casing you must remove it by rubbing with your fingertips To fit a new plate center it with the corners in the space provided and press gently ov...

Page 20: ...s is the numerical position of the digital potentiometer 9 1 3 Speed calibration Testing and setting Select and firm the SPEED CAL 2 submenus and the speed to calibrate The following messages then app...

Page 21: ...requency of controls required based on normal use of the electrocardiograph about 4000 ECG recordings per year Warning Check immediately if there are events not referring to normal use General and saf...

Page 22: ...thermal head correctly use a cloth slightly dampened with alcohol or an equivalent solution with the device off Proceed as follows open the paper feed pass the cloth over the dots of the thermal head...

Page 23: ...R 1200 VIEW 69701441 DISPLAY GLASS FOR AR 1200 VIEW 69701492 ELECTRONIC KEYBOARD BOARD AR 1200 VIEW EXT 69701502 MOTHER BOARD IRDA AR1200VIEW EXT C LINE 69701502E MOTHER BOARD IRDA AR1200VIEW EXT C LI...

Page 24: ...s the insulation required for connection with medical systems The communication with a PC occurs through a proprietary protocol a standard IRDA protocol cannot be used et medical devices has validated...

Page 25: ...ime constant 3 3 s Acquisition 11 bit 1000 samples s channel printing and filters 500 samples s channel in calculation and filters Resolution 5 V bit Leads 9 9 9 0 9 8 acquired 4 derived III aVR aVL a...

Page 26: ...g in real time Automatic simultaneous acquisition ECG Autotimer acquisition at user defined intervals Paper Saving signal acquisition in simultaneous mode with possibility of copying or storing on PC...

Page 27: ...storage Environment temperature from 10 C to 40 C Relative humidity from 10 to 95 without condensation Atmospheric pressure from 500 to 1060 hPa Dimensions 320 x 72 x 240 mm length x height x depth W...

Page 28: ...onductive material and earthed The person performing the repair must be earthed with a suitable protective bracelet compliant with safety regulations The material must be contained in suitable antista...

Page 29: ...29 AR1200VIEW CARDIOLINE is an et medical devices SpA brand w w w c a r d i o l i n e b i z 16 APPENDIX B 16 1 Figures and illustrated tables 0 D...

Page 30: ...DE VIEW VISTA LATERALE SIDE VIEW INGRESSO PAZIENTE PATIENT INPUT TASTO RESET RESET KEY 1 1 2 3 4 BATTE RY PACK 1 2 VITI FISSAGGIO FONDO MOBILE LOWER HOUSING FIXING SCREWS VITE FISSAGGIO SPORTELLO VANO...

Page 31: ...ORE TESTA TERMICA THERMAL HEAD FLAT CONNECTOR FLAT CONNETTORE DISPLAY DISPLAY FLAT CONNECTOR 6 8 7 9 FLAT CONNETTORE TASTIERA KEYBOARD FLAT CONNECTOR INTERFACCIA SERIALE I R SERIAL INTERFACE I R FLAT...

Page 32: ...I FISSAGGIO SCHEDA MADRE MOTHER BOARD FIXING SCREWS 2 1 3 4 5 VITI FISSAGGIO SCHEDA TASTIERA KEYBOARD FIXING SCREWS SCHEDA TAST IERA KEYBOARD VITI FISSAGGIO SISTEMA SCRIVENTE WRITING SYSTEM FIXING SCR...

Page 33: ...COLARI DETAILS VIEW CHIMOGRAFO MOTOR ASSEMBLY T3 FLAT STAMPANTE FLAT THERMAL HEAD SPORTELLO CARTA PAPER FEED SISTEMA SCRIVENTE WRITING SYSTEM RULLO GUIDA CARTA PAPER GUIDE ROLLER MOZZO PORTA CARTA PAP...

Page 34: ...a testa termica ruotando in senso orario o antiorario l apposito ingranaggio max 1 giro dell attrezzo Nel solo caso di stampa non uniforme procedere come segue WRITER MOUNTING INSTRUCTIONS Note Remove...

Page 35: ...N 22 e 23 di J49 STROBE 1 e 2 STROBE 3 e 4 PIN 7 DI J49 LATCH tsetup LATCH thold LATCH thold DATA tsetup DATA 333 nSEC DATA PIN 6 DI J49 CLOCK 3MHz 1 msec TEST POINT TESTA TERMICA THERMAL HEAD TEST PO...

Page 36: ...DIOLINE is an et medical devices SpA brand w w w c a r d i o l i n e b i z 17 Revision Sheet Revision Sheet Service Manual ECG AR 1200 VIEW Lang English CODE 80509531 80509532 REV OM DESCRIPTION DATE...

Page 37: ...an et medical devices SpA brand w w w c a r d i o l i n e b i z DIAGRAMS LIST ECG ar 1200 view CARDIOLINE 115 V 115 V 50 60 Hz Code 80509531 80509532 3 0 1 0 0 1 0 2 222 77777777777777 222 222 222 222...

Page 38: ...38 AR1200VIEW CARDIOLINE is an et medical devices SpA brand w w w c a r d i o l i n e b i z...

Reviews: