7 Trouble Shooting
Defect
Possible cause/symptom
Remedy
The device does not switch on in any
mode.
Mother board/keyboard card.
Holding down the ON/OFF key,
measure 12Vdc on J52 (keyboard
connector) pin 3; if absent replace the
keyboard card, if not, replace the
motherboard.
The device does not switch on when
in mains mode.
Mains LED indicator off.
Check mains fuses, if OK replace
motherboard.
The device does not switch on when
in battery mode.
Check T5A fuse on motherboard.
Check efficiency of battery and
replace if necessary.
The battery does not charge.
The device has not been charged for
long enough.
Mains LED indicator off.
Defective battery.
Leave the device charging for at least
24 hours.
Check mains fuses, if OK replace
motherboard.
Check efficiency of battery and
replace if necessary.
Check T5A fuse on motherboard.
Keyboard keys not working..
The key pressed not enabled for the
specific function.
Execute keyboard self test and if it
fails replace keyboard card.
The paper feeds through without
printing.
Paper not suitable, without black page
recognition mark.
Check that original paper is being
used.
Check the paper has been inserted
correctly.
Check that the 26 Vdc VTPH power
supply to the thermal print head is
present when printing is active, if not,
replace the mother board.
Check signals on test point table T8
are present. If good, replace thermal
head unit, otherwise replace
motherboard.
Anomalous printout.
Clean thermal head.
Execute printer self-test, replace
thermal head unit if some dots are
missing.
Does not print automatically or does
not paginate the trace correctly.
Paper not suitable, without black page
recognition mark.
Clean mark sensor
Perform mark sensor calibration.