Cardoline 80509531 Technical & Service Manual Download Page 14

14 

AR1200VIEW

 

CARDIOLINE

®

 is an et medical devices SpA brand 

w w w . c a r d i o l i n e . b i z

Do not dispose of a spent battery as ordinary refuse or litter. 

Note: The operation of the device is not guaranteed if the battery is flat or missing. 

Note: The configuration parameters and any stored ECGs are not lost when the battery is changed. 

Note: Update the system date and time after replacing the battery. 

Battery check 

Proceed as follows to check the efficiency of the battery: 

  Leave the battery on charge for at least 18 hours; 

  Disconnect the mains cable; 

  Activate printing of an ECG in manual mode on 3 channels at 5 mm/s: if the device shows the battery is flat within the first 5 

minutes the battery should be replaced. 

6.4  Keyboard card 

The keyboard consists of: 

  18 function keys; 

  10 alphanumeric keys; 

  1 LED for mains line indication; 

  CPU connection through bi-directional synchronous serial port. 

Testing the keyboard card circuits 

  Execution of keyboard self-test; 

6.5  Printer unit 

This consists of the thermal printer head support and the mechanical elements required to position it correctly. 

  864 dot high resolution printer head, 8 dots per mm; 

  support; 

  paper feed unit consisting of 1 direct current reduction gears, equipped with stroboscopic speed control; 

  mark detection sensor. 

Testing the print unit 

  Execution of printer self-test. 

6.6  IR Module 

 

The function of the IR module is to transmit and receive data from an external PC. The IR module can be used to perform the 

following functions:

 

 

Updating firmware (Loader)*;

 

 

Batch ECG transmission (L.E.M.S.)*;

 

  Real time ECG transmission (R.T.E ir)*. 

 
*Consult the specific manual for further information. 

 

 
 
 
 
 
 
 

Summary of Contents for 80509531

Page 1: ...w w c a r d i o l i n e b i z T TE EC CH HN NI IC CA AL L S SE ER RV VI IC CE E M MA AN NU UA AL L E EL LE EC CT TR RO OC CA AR RD DI IO OG GR RA AP PH H M MO OD DE EL L C Ca ar rd di io ol li in ne...

Page 2: ...2 AR1200VIEW CARDIOLINE is an et medical devices SpA brand w w w c a r d i o l i n e b i z Index 0 1 2 3 4 2 3 0 1 4 5 1 6...

Page 3: ...3 AR1200VIEW CARDIOLINE is an et medical devices SpA brand w w w c a r d i o l i n e b i z 2 1 7 2 2 2 3 2 2 2 2 2 4 8 9 0 9 0 9 0 3...

Page 4: ...o store up to 40 full ECG exams with no need to print out immediately on paper paper saving mode ECG measurements option automatic ECG parameter measurement program ECG signal interpretive option a us...

Page 5: ...that the signal is restored use only original electrodes or electrodes responding to IEC and AAMI standards If an electrosurgical scalpel is in use the patient cable should be disconnected from the d...

Page 6: ...MH 12V 2 3 Display Display assembly is composed of frame monochromatic LCD display 320 x 240 pixel 4 7 inch as well as a step up converter 2 4 Keyboard The keyboard card consists of the device functio...

Page 7: ...to the patient input port 8 4 1 2 3 6 7 5 15 9 10 11 12 13 14 Port from connection side Pin 1 IN C2 electrode C2 Pin 2 IN C3 electrode C3 Pin 3 IN C4 electrode C4 Pin 4 IN C5 electrode C5 Pin 5 IN C6...

Page 8: ...nstrument s manual recalling that The leakage current to the casing is measuring between the mains supply circuits and a metal sheet no greater than 20 x 10 cm pressed against the casing of the device...

Page 9: ...nnected to the device to the positive pin of instrument 5 1 a connect all the other cable terminals to the negative pin of instrument 5 1 a record the signal for a few seconds check that the amplitude...

Page 10: ...the signal on leads V1 V6 measure the length of the wave cycle recorded on the paper The results should be as follows Period 50 mm 5 for a feed rate of 50 mm s Period 25 mm 5 for a feed rate of 25 mm...

Page 11: ...3 2 0 0 03 4 2 03 4 3 2 2 2C 3 2 22 2C 7 2C 28 3 22 2 2C 7 2C Note The 0 5 Hz pitch linear phase anti drift filter is always on and cannot be switched off The 50 or 60 Hz filter eliminates modified no...

Page 12: ...illustrated in the picture below Changing mains voltage does not require mains fuses rate modification 5 Warning Changing mains voltage imply the replacement of the mentioned label Only original ones...

Page 13: ...nal polarisation circuit Hybrid circuit for amplification filter and clamp 6 1 6 Motor control This consists of the following parts Stroboscopic sensor control Phase comparator between reference frequ...

Page 14: ...d 6 4 Keyboard card The keyboard consists of 18 function keys 10 alphanumeric keys 1 LED for mains line indication CPU connection through bi directional synchronous serial port Testing the keyboard ca...

Page 15: ...15 AR1200VIEW CARDIOLINE is an et medical devices SpA brand w w w c a r d i o l i n e b i z...

Page 16: ...mains fuses if OK replace motherboard Check efficiency of battery and replace if necessary Check T5A fuse on motherboard Keyboard keys not working The key pressed not enabled for the specific function...

Page 17: ...nd display assembly Verify on display connector J49 presence of LCD data signal level from 0 trough 5 V if present replace display assembly otherwise replace mother board 8 How to dismantle and reasse...

Page 18: ...sembly proceed in inverse order 8 8 Removing the printer assembly table T2b Proceed as follows proceed as in chapter 8 3 for opening the equipment and removing the mother board from the equipment remo...

Page 19: ...pull it off the casing If any adhesive is left on the casing you must remove it by rubbing with your fingertips To fit a new plate center it with the corners in the space provided and press gently ov...

Page 20: ...s is the numerical position of the digital potentiometer 9 1 3 Speed calibration Testing and setting Select and firm the SPEED CAL 2 submenus and the speed to calibrate The following messages then app...

Page 21: ...requency of controls required based on normal use of the electrocardiograph about 4000 ECG recordings per year Warning Check immediately if there are events not referring to normal use General and saf...

Page 22: ...thermal head correctly use a cloth slightly dampened with alcohol or an equivalent solution with the device off Proceed as follows open the paper feed pass the cloth over the dots of the thermal head...

Page 23: ...R 1200 VIEW 69701441 DISPLAY GLASS FOR AR 1200 VIEW 69701492 ELECTRONIC KEYBOARD BOARD AR 1200 VIEW EXT 69701502 MOTHER BOARD IRDA AR1200VIEW EXT C LINE 69701502E MOTHER BOARD IRDA AR1200VIEW EXT C LI...

Page 24: ...s the insulation required for connection with medical systems The communication with a PC occurs through a proprietary protocol a standard IRDA protocol cannot be used et medical devices has validated...

Page 25: ...ime constant 3 3 s Acquisition 11 bit 1000 samples s channel printing and filters 500 samples s channel in calculation and filters Resolution 5 V bit Leads 9 9 9 0 9 8 acquired 4 derived III aVR aVL a...

Page 26: ...g in real time Automatic simultaneous acquisition ECG Autotimer acquisition at user defined intervals Paper Saving signal acquisition in simultaneous mode with possibility of copying or storing on PC...

Page 27: ...storage Environment temperature from 10 C to 40 C Relative humidity from 10 to 95 without condensation Atmospheric pressure from 500 to 1060 hPa Dimensions 320 x 72 x 240 mm length x height x depth W...

Page 28: ...onductive material and earthed The person performing the repair must be earthed with a suitable protective bracelet compliant with safety regulations The material must be contained in suitable antista...

Page 29: ...29 AR1200VIEW CARDIOLINE is an et medical devices SpA brand w w w c a r d i o l i n e b i z 16 APPENDIX B 16 1 Figures and illustrated tables 0 D...

Page 30: ...DE VIEW VISTA LATERALE SIDE VIEW INGRESSO PAZIENTE PATIENT INPUT TASTO RESET RESET KEY 1 1 2 3 4 BATTE RY PACK 1 2 VITI FISSAGGIO FONDO MOBILE LOWER HOUSING FIXING SCREWS VITE FISSAGGIO SPORTELLO VANO...

Page 31: ...ORE TESTA TERMICA THERMAL HEAD FLAT CONNECTOR FLAT CONNETTORE DISPLAY DISPLAY FLAT CONNECTOR 6 8 7 9 FLAT CONNETTORE TASTIERA KEYBOARD FLAT CONNECTOR INTERFACCIA SERIALE I R SERIAL INTERFACE I R FLAT...

Page 32: ...I FISSAGGIO SCHEDA MADRE MOTHER BOARD FIXING SCREWS 2 1 3 4 5 VITI FISSAGGIO SCHEDA TASTIERA KEYBOARD FIXING SCREWS SCHEDA TAST IERA KEYBOARD VITI FISSAGGIO SISTEMA SCRIVENTE WRITING SYSTEM FIXING SCR...

Page 33: ...COLARI DETAILS VIEW CHIMOGRAFO MOTOR ASSEMBLY T3 FLAT STAMPANTE FLAT THERMAL HEAD SPORTELLO CARTA PAPER FEED SISTEMA SCRIVENTE WRITING SYSTEM RULLO GUIDA CARTA PAPER GUIDE ROLLER MOZZO PORTA CARTA PAP...

Page 34: ...a testa termica ruotando in senso orario o antiorario l apposito ingranaggio max 1 giro dell attrezzo Nel solo caso di stampa non uniforme procedere come segue WRITER MOUNTING INSTRUCTIONS Note Remove...

Page 35: ...N 22 e 23 di J49 STROBE 1 e 2 STROBE 3 e 4 PIN 7 DI J49 LATCH tsetup LATCH thold LATCH thold DATA tsetup DATA 333 nSEC DATA PIN 6 DI J49 CLOCK 3MHz 1 msec TEST POINT TESTA TERMICA THERMAL HEAD TEST PO...

Page 36: ...DIOLINE is an et medical devices SpA brand w w w c a r d i o l i n e b i z 17 Revision Sheet Revision Sheet Service Manual ECG AR 1200 VIEW Lang English CODE 80509531 80509532 REV OM DESCRIPTION DATE...

Page 37: ...an et medical devices SpA brand w w w c a r d i o l i n e b i z DIAGRAMS LIST ECG ar 1200 view CARDIOLINE 115 V 115 V 50 60 Hz Code 80509531 80509532 3 0 1 0 0 1 0 2 222 77777777777777 222 222 222 222...

Page 38: ...38 AR1200VIEW CARDIOLINE is an et medical devices SpA brand w w w c a r d i o l i n e b i z...

Reviews: