Do not dispose of a spent battery as ordinary refuse or litter.
Note: The operation of the device is not guaranteed if the battery is flat or missing.
Note: The configuration parameters and any stored ECGs are not lost when the battery is changed.
Note: Update the system date and time after replacing the battery.
Battery check
Proceed as follows to check the efficiency of the battery:
Leave the battery on charge for at least 18 hours;
Disconnect the mains cable;
Activate printing of an ECG in manual mode on 3 channels at 5 mm/s: if the device shows the battery is flat within the first 5
minutes the battery should be replaced.
6.4 Keyboard card
The keyboard consists of:
18 function keys;
10 alphanumeric keys;
1 LED for mains line indication;
CPU connection through bi-directional synchronous serial port.
Testing the keyboard card circuits
Execution of keyboard self-test;
6.5 Printer unit
This consists of the thermal printer head support and the mechanical elements required to position it correctly.
864 dot high resolution printer head, 8 dots per mm;
support;
paper feed unit consisting of 1 direct current reduction gears, equipped with stroboscopic speed control;
mark detection sensor.
Testing the print unit
Execution of printer self-test.
6.6 IR Module
The function of the IR module is to transmit and receive data from an external PC. The IR module can be used to perform the
following functions:
Updating firmware (Loader)*;
Batch ECG transmission (L.E.M.S.)*;
Real time ECG transmission (R.T.E ir)*.
*Consult the specific manual for further information.