10
7.3.3 Power Inlets
The chromatography refrigerator has two power inlets at the back of the unit:
1.
Main Power Inlet (A)
- This is used to connect the AC mains power cord to the chromatography refrigerator.
2.
Chromatography Power Inlet (C)
- This is used to connect the AC mains power cord to the chromatography outlet
assembly inside of the cabinet.
7.3.4 Switches
The chromatography refrigerator has three switches at the back of the unit:
1.
Mains Power Inlet Switch (B)
- This is used to turn ON or OFF the AC supply to the refrigerator.
2.
Chromatography Power Inlet Switch (D)
- This is used to turn ON or OFF the AC inlet supply to the receptacle
inside the refrigerator.
3.
Chromatography Manual Reset Switch (E)
- This is used to manually reset the safety circuit in the AC receptacle
assembly after the LED turns RED from GREEN due to a safety trip event which disengages power to outlet inside the
refrigerator.
Figure 2. Chromatography Refrigerator Power Inlet and Switches
Refer to the label adjacent to the switches at the back of the refrigerator for identification.