8
7. Installation
WARNING:
Do not exceed the electrical rating printed on the dataplate located on the upper left side of the unit.
CAUTION:
Do not move the unit using the drain pan on the back. This could cause damage to the equipment.
7.1 Location
Install the unit on a level area free from vibration with a minimum of 6 inches of space on the sides and rear and
12 inches at the top. Do not position the equipment in direct sunlight or near heating diffusers, radiators,
or other sources of heat.
WARNING:
Do not move the unit while loaded. Unit shall always be moved when empty and by pushing slowly at
handle level or lower on the unit. Be especially careful on any uneven surfaces.
7.1.1 Installation Instructions
The unit must be level both front to back and side to side when installed. If the unit is out of level, you may need to shim
the corners or casters with thin sheets of metal. Be sure to set the brakes for units equipped with casters.
CAUTION:
An unlevel unit may result in instability and performance issues for the doors and drawers.
7.2 Wiring
Wiring diagrams are attached on the back of the cabinet.
CAUTION:
Connect the equipment to the correct power source. Incorrect voltage can result in severe damage
to the equipment.
CAUTION:
For personal safety and trouble-free operation, this unit must be properly grounded while in use.
Failure to ground the equipment may cause personal injury or damage to the equipment. Always conform
to the National Electrical Code and local codes. Do not connect the unit to overloaded power lines.
CAUTION:
Do not position the unit in a way that impedes access to the disconnecting device or
circuit breaker in the back of the unit.
CAUTION:
Always connect the unit to a dedicated (separate) circuit. Each unit is equipped with a service cord and plug
designed to connect it to a power outlet which delivers the correct voltage. Supply voltage must be within ±10% of the
unit rated voltage. If cord becomes damaged, replace with a properly rated power supply cord
Table 3. Power Cord Specifications
CAUTION:
Never cut the grounding prong from the service cord plug. If the prong is removed,
the warranty is invalidated.
CAUTION:
In an emergency, the power cord is a disconnect device.
Model
Power Cord Specifications
A
3-G 12 AWG, NEMA 5-15P, 15A/125V
D
3-G 14 AWG, NEMA 6-15P, 15A/250V