General Information
2-75
Alaris
®
System Directions for Use
Pump and Syringe Modules Section
Before operating instrument, verify that administration set is
free from kinks and installed correctly
in instrument.
CAUTION
•
Epidural administration
of drugs other than those
indicated for epidural use could result in serious injury to
the patient.
• It is strongly recommended that the source container,
administration set, and Pump Module used for
epidural
drug delivery
be clearly differentiated from those used
for other types of administration.
• It is strongly recommended that the syringe,
administration set and Syringe Module used for
epidural drug delivery
be clearly differentiated from
those used for other types of administration.
WARNINGS
Epidural Administration
• When the
pressure sensing disc is not being used
and an occlusion occurs, there is a risk of infusing
pressurized buildup of infusates upon correction of the
occlusion. To avoid an inadvertent bolus, relieve the
pressure before restarting the infusion.
• Ensure the displayed
syringe manufacturer and
syringe size
correctly identify the installed syringe.
Mismatches may cause an under-infusion or over-
infusion to the patient that could result in serious injury
and/or death. For a list of compatible syringes,
reference "Compatible Syringes".
•
Installing a pressure sensing disc
after an infusion
has started can result in a bolus to the patient.
WARNINGS
Administration Sets
(Continued)
Warnings and Cautions
(Continued)
ALARM
INFUSE
STANDBY
CHANNEL
SELECT
CHANNEL
OFF
PAUSE
RESTART
RATE (mL/h)
ALARM
INFUSE
STANDBY
CHANNEL
SELECT
CHANNEL
OFF
PAUSE
RESTART
RATE (mL/h)
CHANNEL
SELECT
CHANNEL
OFF
RESTART
PAUSE
ALARM
INFUSE
STANDBY
ALARM
INFUSE
STANDBY
CHANNEL
SELECT
CHANNEL
OFF
PAUSE
RESTART
RATE (mL/h)
ALARM
INFUSE
STANDBY
CHANNEL
SELECT
CHANNEL
OFF
PAUSE
RESTART
RATE (mL/h)
ALARM
INFUSE
STANDBY
CHANNEL
SELECT
CHANNEL
OFF
PAUSE
RESTART
RATE (mL/h)
CHANNEL
SELECT
CHANNEL
OFF
RESTART
PAUSE
ALARM
INFUSE
STANDBY
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......