Oxygen Calibration Alarm:
An audible and visual Oxygen Calibration
alarm will indicate if an oxygen sensor has failed to calibrate. If this alarm
occurs, remove patient and cycle the power for re-calibration or remove the
driver from service. While the alarm is active, the driver provides CPAP and
FiO
2
, however, the FiO
2
may not be accurate. The system calibrates the
oxygen sensor during start up and routinely during operation. There is no
delay for this alarm
Software Fault:
An audible and visual Software Fault alarm will indicate
any software fault detected. The system will display a fault code which
should be recorded. If the fault is non-fatal, the alarm can be cleared and
the system will return to normal operation. If the fault is fatal, the driver will
need to have the power cycled or be removed from service. There is no
delay for this alarm
Flow / Pressure Sensor:
A visual Sensor Calibration Fault will indicate if a
sensor has failed the calibration during the start up routine. A failure screen
will appear indicating what sensor has failed and informing the user to
remove the device from service.
IEC 60601-1 Classification
Internally Powered Equipment, Type BF, IPX 1
This equipment is rated for Continuous Operation.
Caution: Although the Air
Life
®
nCPAP System Driver meets
the requirements of current EMC/RFI legislation, this does not
guarantee immunity from all sources of radiated energy. Some
mobile telephones and other products containing radio
transmitting components may cause a malfunction of the
Air
Life
®
nCPAP driver and should not be used in the vicinity of
the driver.
Storage and Operating Conditions
•
Safe for storage at -20°C–60°C
•
Relative humidity of 15%–95% non-condensing
•
Atmospheric pressure 0.6–1.4 Bar.
•
Operating environment temperature is 5°C–40°C
•
Keep dry and do not expose to direct sunlight
•
This product meets the EMC requirements of IEC60601-1
Warning: The patient accessories for use with the
Air
Life
®
nCPAP Driver are for single patient use only. Disposable
accessories include generators, fixation device, prongs,
masks, heated wire breathing circuits, and humidification
chambers. DO NOT STERILIZE OR REUSE.
6 Air
Life
®
nCPAP Operator’s Manual