Caution: The driver will be damaged if any part of it is immersed in liquid or
gas, or steam sterilized.
Note:
Alarms are delayed for 120 seconds after entering the main screen in
order to allow time for patient setup. If the patient setup takes longer than
120 seconds, the alarms can be muted for an additional 120 seconds by
pressing the alarm mute/reset button or the system can be placed into
Standby Mode.
Note:
To enhance patient safety and comfort, use only with heated
humidifiers and heated wire circuits. The recommended setting is 37 °C,
100% RH.
Note:
To obtain maximum battery life, it is recommended that the driver
remain connected to an external power source even when it is not in use.
Note:
Class 1 LED product.
Note:
Hospital personnel should keep records of Performance Verification
Tests and identify equipment that is authorized for use.
Note:
The Driver must have both gases properly connected and within the
specified pressure range in order to pass the initial start-up sequence. This
will enable the Driver to perform a valid calibration and ensure accurate FiO
2
delivery.
Air
Life
®
nCPAP Operator’s Manual
11