70-00297-01 A
TM 55/65 Treadmill 4.0
1-12
Safety Requirements
Power
frequency (50/
60 Hz)
Magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment
Note:
U
T
is the A.C. mains voltage prior to application of the test level.
Note:
Tests verified with shielded input/output cables only.
Table 1-3: Guidance and Manufacturer's Declaration - Electromagnetic Immunity
Immunity Test
IEC 60601
test level
Compliance
Level
Electromagnetic Environment -
Guidance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to
80 MHz
3 V
Portable and mobile RF
communications equipment
should be used no closer to any
part of the ST/TM55 or ST/TM65,
including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation
distance:
150 kHz to 80 MHz:
Table 1-2: Guidance and Manufacturer's Declaration - Electromagnetic Immunity
Immunity Test
IEC 60601
test level
Compliance
Level
Electromagnetic
Environment - Guidance
d
1.2
P
=
Summary of Contents for Cardiac Stress
Page 1: ...Cardiac Stress Treadmills User s Guide 70 00297 01 A...
Page 2: ...CARDIAC STRESS TREADMILLS USER S GUIDE 70 00297 01 A...
Page 25: ...70 00297 01 A TM 55 65 Treadmill 4 0 2 6 Introduction...
Page 31: ...70 00297 01 A TM 55 65 Treadmill 4 0 3 6 Operating the Treadmill...
Page 45: ...70 00297 01 A TM 55 65 Treadmill 4 0 4 14 Maintenance Troubleshooting...
Page 51: ...70 00297 01 A TM 55 65 Treadmill 4 0 A 6...