70-00297-01 A
TM 55/65 Treadmill 4.0
B-6
Connecting to the Monitoring System
To connect the treadmill to the monitoring system:
1.
Be sure the treadmill and the monitoring system are disconnected from the power
source.
2.
Inspect the monitoring system interface cable connector, located on the front of the
hood (see
Figure B-2
). If it appears damaged, contact an authorized technician before
continuing this installation.
If the connector is undamaged, plug the male end of the monitoring system interface
cable into the connector. Tighten the connector thumb screws until finger tight.
3.
Plug the female end into the treadmill connector on the back panel of the monitoring
system.
4.
Verify that the outlet voltage matches the voltage on the nameplate located on the
front of the treadmill hood.
5.
If the voltage is adequate, plug in the power cord. Do not use this outlet if the voltage
is not adequate.
Figure B-2:
Example of Monitoring System Connections/
1
Monitoring System Interface Cable Connector
2
Emergency Stop Switch Connector
3
Circuit Breaker Switch
4
USB Monitoring System Interface Cable Connector
3
4
2
1