70-00297-01 A
TM 55/65 Treadmill 4.0
1-10
Safety Requirements
EMC Declaration Tables
This section lists EMC Declaration Tables.
The Cardiac Stress Treadmill requires special precautions regarding EMC and must be
installed and put into service according to the guidelines of the EMC declaration
tables.
Portable and mobile RF communications equipment may affect the Cardiac Stress
Treadmill and the recommended separation distances in the EMC declaration tables
should be observed.
Electromagnetic Emissions
Cardiac Stress Treadmills are intended for use in the electromagnetic environment
specified below. The customer or user of this equipment must ensure it is used in such an
environment.
Table 1-1: Guidance and Manufacturer's Declaration - Electromagnetic Emissions
Emissions Test
Compliance
Electromagnetic Environment -
Guidance
RF emissions
CISPR 11
Group 1
The Cardiac Stress Treadmills use RF
energy only for its internal function.
Therefore, its RF emissions are very low
and are not likely to cause any
interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class A
The Cardiac Stress Treadmills are
suitable for use in all establishments
other than domestic and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
Note:
Tests verified with shielded input/output cables only.
Summary of Contents for Cardiac Stress
Page 1: ...Cardiac Stress Treadmills User s Guide 70 00297 01 A...
Page 2: ...CARDIAC STRESS TREADMILLS USER S GUIDE 70 00297 01 A...
Page 25: ...70 00297 01 A TM 55 65 Treadmill 4 0 2 6 Introduction...
Page 31: ...70 00297 01 A TM 55 65 Treadmill 4 0 3 6 Operating the Treadmill...
Page 45: ...70 00297 01 A TM 55 65 Treadmill 4 0 4 14 Maintenance Troubleshooting...
Page 51: ...70 00297 01 A TM 55 65 Treadmill 4 0 A 6...