![Capintec CRC-712MX Owner'S Manual Download Page 43](http://html.mh-extra.com/html/capintec/crc-712mx/crc-712mx_owners-manual_494184043.webp)
CAPINTEC, INC.
CRC
®
-712MX
The CRC
®
-712MX can be set to operate from either a nominal 115 volt or a nominal 230 volt
AC power line. To check or change the setting: (refer to Figure 7-1)
1. Turn off the CRC
®
-712MX power switch and unplug the line cord from the power
entry module.
2. Slide the clear plastic door to the left.
3. Pull the FUSE PULL lever and remove the fuse.
4. The number which is visible is the present setting.
5. To change the setting, use a small pair of pliers or a wire bent into a hook and
remove the card. Rotate the card so that the desired voltage will be readable, and
push the card back into place.
6. Install a fuse with the type and rating that is specified for the selected line voltage and
slide the clear plastic door to the right.
a. For the 115 volt power line, use a 1/4 Amp slow-blow type fuse.
b. For the 230 volt power line, use a 1/8 Amp slow-blow type fuse.
7. Install a power cord that is specified for the selected line voltage.
a. For the 115 volt power line, use a detachable power cord with a NEMA type 5-
15R 3 prong plug similar to Belden 17250.
b. For the 230 volt power line, use a detachable power cord with a NEMA type 6-
15R 3 prong plug similar to Belden 17566.
8. Verify the CRC
®
-712MX System is functioning correctly by performing the Daily Test
as specified in CHAPTER 5: QUALITY ASSURANCE & ACCEPTANCE TESTING.
Figure 7-1
Printer Fuse
The printer fuse is not accessible from the outside of the printer case and must be replaced
by a qualified service representative.
January 08
CLEANING AND MAINTENANCE
7 - 3