CAPINTEC, INC.
CRC
®
-712MX
Shielding:
Personnel Protection Shield of 3.2mm (1/8in) mechanically
reinforced high density lead shielding; with 1Ci of Tc99m in the
chamber, the exposure 50cm in front of the instrument is less
than 0.5mR/min.
Overall Accuracy
Overall accuracy of the calibrator is determined by the accuracy of:
•
specific source calibration
•
chamber linearity and response
•
iometer
•
digital
readout
Repeatability
Measurements will repeat to within, ±1% for a period of 24 hours during which time
the instrument is maintained under constant temperature, humidity, and background
radiation conditions, and is powered at all times.
Fig, 2-1 CRC
®
-712MX Simplified Block Diagram
2 - 6
FUCTIONAL & TECHNICAL DESCRIPTION
January 08