Intercomparison Calibration with a 12 Atmosphere Chamber
1. Verify that the Zero and Background adjustments of the calibrator are within the
specified limits. (Reference CHAPTER 5: QUALITY ASSURANCE AND
ACCEPTANCE TESTING; SECTION: ACCEPTANCE TESTING.)
2. Insert a sample of the radioisotope, with adequate activity to display 100µCi or higher,
into the chamber of the 12 atmosphere calibrator.
3. Select
the
OTHER
push-button and adjust the Calibration Potentiometer to the value
listed in Table II of the owner’s manual for a standard CRC
®
Calibrator.
4. Record the displayed activity for the measured source.
5. Place the SAME source in the CRC
®
-712MX chamber.
6. Select
the
OTHER
push-button and adjust the Calibration Control to obtain the same
displayed activity as was recorded in step 4.
7. Record the number appearing on the Calibration Control knob-dial. This number may
now be used for all future measurements of the radioisotope of similar sample
configuration