CAPINTEC, INC.
CAPTUS
®
600
checking its system memory. A display similar to Figure 6-19 will appear (the PROM values
shown are just an example). The same results will also be sent to the printer.
1 4 :3 2
D IA G N O S T IC S
R A M : P A S S
P R O M : P A S S
E E P R O M : P A S S
9B 8 7 72 2 9 1 E B E
80 E F 38 4 0 2 D 2 C
E 7 10 7 89 2 5 7 F 6
P re ss a n y k e y to c o n tin u e
Figure 6-19
If any of the test results indicate “FAIL” rather than “PASS”, it indicates a system malfunction.
If this happens, perform the Diagnostics Utility again. If the test(s) fail again, contact
Capintec’s Authorized Serve Center for assistance.
CAUTION!
Do NOT continue to use the CAPTUS
®
600
if any Diagnostics test results read “FAIL”.
Diagnostics Printout
On the following pages is a sample Diagnostics printout. Below is a description of the
sections in the Diagnostics
printout.
System Configuration
This data is entered into the system memory at the factory. It indicates to the
CAPTUS
®
600 which detectors are connected on the back panel. The user cannot
change this.
Nuclide Data
Nuclides List
A listing of Nuclides in the CAPTUS
®
600 memory. This list is entered at the factory.
Additional nuclides may be added to this list.
Half-Life
A listing of the Half-life for each nuclide in the CAPTUS
®
600 memory.
Photon Energies
A listing of the Energy Photon Peaks used by the CAPTUS
®
600 in the automatic
peak searching algorithm.
November 07
ACCEPTANCE & QUALITY ASSURANCE TESTS
6-13
Summary of Contents for CAPTUS 600
Page 2: ......
Page 40: ......
Page 62: ......
Page 77: ...CAPINTEC INC CAPTUS 600 Page 1 November 07 ACCEPTANCE QUALITY ASSURANCE TESTS 6 15...
Page 78: ...CAPINTEC INC CAPTUS 600 Page 2 Page 3 6 16 ACCEPTANCE QUALITY ASSURANCE TESTS November 07...
Page 90: ......
Page 102: ......
Page 112: ...CAPINTEC INC CAPTUS 600 Figure 9 15 9 10 THYROID UPTAKE ONE CAPSULE METHOD November 07...
Page 120: ...CAPINTEC INC CAPTUS 600 Figure 10 11 10 8 THYROID UPTAKE TWO CAPSULES METHOD November 07...
Page 131: ...CAPINTEC INC CAPTUS 600 Figure 11 16 November 07 THYROID UPTAKE STANDARD CAPSULE METHOD 11 11...
Page 132: ......
Page 150: ...CAPINTEC INC CAPTUS 600 Figure 13 11 13 8 WIPE TEST November 07...
Page 152: ......
Page 162: ...CAPINTEC INC CAPTUS 600 Figure 15 5 15 4 LAB TESTS URINE TEST November 07...
Page 167: ...CAPINTEC INC CAPTUS 600 Figure 16 5 November 07 LAB TESTS BLOOD TESTS 16 5...
Page 171: ...CAPINTEC INC CAPTUS 600 Figure 16 9 November 07 LAB TESTS BLOOD TESTS 16 9...
Page 175: ...CAPINTEC INC CAPTUS 600 Figure 16 14 November 07 LAB TESTS BLOOD TESTS 16 13...
Page 176: ......
Page 186: ......
Page 208: ......
Page 214: ......
Page 216: ......
Page 217: ......
Page 218: ......