CAPINTEC, INC.
CAPTUS
®
600
CHAPTER 4
SYSTEM SETUP
INSTALLATION
Your system has been shipped from the factory pre-configured and fully tested. Installation
consists of placing the components on the stand and securing them in place as shown in the
CAPTUS
®
600 Unpacking and Assembly Instructions
document.
ENVIRONMENT REQUIREMENTS
Operational
The instrument should be located where the level of the background radiation is as low and
as constant as possible.
The instrument should be located where the temperature is stable within a range of +50°F to
+85°F (+10°C to +30°C) and the maximum relative humidity is 90% non-condensing to
warrant maximum reliability and accuracy.
The instrument should be located where the barometric pressure is within a range of 27–31
inches of mercury (91–105 kilopascals).
Storage
The instrument should be stored where the temperature is stable and the range is from
+39°F to +110°F (+4°C to +43°C) and the maximum relative humidity is 90% non-condensing
to warrant maximum reliability.
The instrument should be stored where the barometric pressure is within a range of 15–33
inches of mercury (51–112 kilopascals).
CAUTION
If these environmental requirements are not followed,
the instrument may display erroneous readings.
November 07
SYSTEM SETUP
4-1
Summary of Contents for CAPTUS 600
Page 2: ......
Page 40: ......
Page 62: ......
Page 77: ...CAPINTEC INC CAPTUS 600 Page 1 November 07 ACCEPTANCE QUALITY ASSURANCE TESTS 6 15...
Page 78: ...CAPINTEC INC CAPTUS 600 Page 2 Page 3 6 16 ACCEPTANCE QUALITY ASSURANCE TESTS November 07...
Page 90: ......
Page 102: ......
Page 112: ...CAPINTEC INC CAPTUS 600 Figure 9 15 9 10 THYROID UPTAKE ONE CAPSULE METHOD November 07...
Page 120: ...CAPINTEC INC CAPTUS 600 Figure 10 11 10 8 THYROID UPTAKE TWO CAPSULES METHOD November 07...
Page 131: ...CAPINTEC INC CAPTUS 600 Figure 11 16 November 07 THYROID UPTAKE STANDARD CAPSULE METHOD 11 11...
Page 132: ......
Page 150: ...CAPINTEC INC CAPTUS 600 Figure 13 11 13 8 WIPE TEST November 07...
Page 152: ......
Page 162: ...CAPINTEC INC CAPTUS 600 Figure 15 5 15 4 LAB TESTS URINE TEST November 07...
Page 167: ...CAPINTEC INC CAPTUS 600 Figure 16 5 November 07 LAB TESTS BLOOD TESTS 16 5...
Page 171: ...CAPINTEC INC CAPTUS 600 Figure 16 9 November 07 LAB TESTS BLOOD TESTS 16 9...
Page 175: ...CAPINTEC INC CAPTUS 600 Figure 16 14 November 07 LAB TESTS BLOOD TESTS 16 13...
Page 176: ......
Page 186: ......
Page 208: ......
Page 214: ......
Page 216: ......
Page 217: ......
Page 218: ......