CAPINTEC, INC.
CAPTUS
®
600
INTERPRETING WIPE RESULTS
The data shown in Figure 13-9 and Figure 13-12 are the results of a measurement with only
Cs137.
NEXT
W-01 HOT LAB SINK
COUNT
GRAPH
P
Trig Level: 2000
CF: 1600 87.75 1.05 kdpm
+
CS137 193.5 4.2 kdpm
+
326 keV 4257. 164. cpm
+
+
+
HIGH
HIGH
A
B
C
200 keV 5086. 179. cpm
376 keV 3953. 158. cpm
Figure 13-12
•
Line A
indicates that the Generic CF result has exceeded the Trigger Level
set up in
the Wipe Setup/Trigger Levels. This is a good indication that either there was
contamination on the wipe, or that the background has changed. If you suspect a
background change, re-measure background and re-measure the wipe.
•
Line B
indicates that the CAPTUS
®
600 has detected a peak within 5% of the energy
of the main peak of Cs137. Although there is a good chance that the CAPTUS
®
600
has identified the presence of the nuclide on the wipe, it has only identified a peak at
a particular energy. The user must decide if the data is consistent with nuclides used
in the department.
•
Line C
indicates that peaks were found at the indicated energies; they are artifacts.
The peak search algorithm used in Wipe is very sensitive. It is not unusual to detect artifact
peaks. You should enter CFs for the nuclides that are used and therefore, may possibly find
on a wipe, but do not enter CFs for nuclides that you do not use.
Many nuclides have energies for more than 1 peak listed (refer to the Diagnostics printout). If
the nuclide has been found, the secondary and tertiary peaks will not be listed. For example,
Cs137's 33 keV peak was not listed.
If a nuclide’s main peak is within 5% of the peaks of a nuclide that is really in the Wipe, the
second nuclide may be listed even though it is not in the Wipe. Therefore, do not enter CF's
for nuclides that you do not use.
November 07
WIPE TEST
13-9
Summary of Contents for CAPTUS 600
Page 2: ......
Page 40: ......
Page 62: ......
Page 77: ...CAPINTEC INC CAPTUS 600 Page 1 November 07 ACCEPTANCE QUALITY ASSURANCE TESTS 6 15...
Page 78: ...CAPINTEC INC CAPTUS 600 Page 2 Page 3 6 16 ACCEPTANCE QUALITY ASSURANCE TESTS November 07...
Page 90: ......
Page 102: ......
Page 112: ...CAPINTEC INC CAPTUS 600 Figure 9 15 9 10 THYROID UPTAKE ONE CAPSULE METHOD November 07...
Page 120: ...CAPINTEC INC CAPTUS 600 Figure 10 11 10 8 THYROID UPTAKE TWO CAPSULES METHOD November 07...
Page 131: ...CAPINTEC INC CAPTUS 600 Figure 11 16 November 07 THYROID UPTAKE STANDARD CAPSULE METHOD 11 11...
Page 132: ......
Page 150: ...CAPINTEC INC CAPTUS 600 Figure 13 11 13 8 WIPE TEST November 07...
Page 152: ......
Page 162: ...CAPINTEC INC CAPTUS 600 Figure 15 5 15 4 LAB TESTS URINE TEST November 07...
Page 167: ...CAPINTEC INC CAPTUS 600 Figure 16 5 November 07 LAB TESTS BLOOD TESTS 16 5...
Page 171: ...CAPINTEC INC CAPTUS 600 Figure 16 9 November 07 LAB TESTS BLOOD TESTS 16 9...
Page 175: ...CAPINTEC INC CAPTUS 600 Figure 16 14 November 07 LAB TESTS BLOOD TESTS 16 13...
Page 176: ......
Page 186: ......
Page 208: ......
Page 214: ......
Page 216: ......
Page 217: ......
Page 218: ......