CAPINTEC, INC.
CAPTUS
®
600
CAUTIONS AND NOTES
CAUTION
Only qualified/trained personnel should operate or service this unit.
CAUTION
If the equipment is used in a manner not specified in this manual, the protection provided by
the equipment may be impaired.
CAUTION
High voltage is present inside the Detectors (up to 1000 Volts). Due to the presence of these
high voltages, opening the covers with the system plugged in may be hazardous. Refer all
servicing to qualified personnel.
CAUTION
No internal adjustments inside the Detector or Readout may be performed by the user within
the conditions of the warranty, except for changing the fuse. Due to the presence of high
voltages, opening the cover with the system plugged in may be hazardous. Refer all
servicing to qualified personnel.
CAUTION
Never use the Well Detector without the well liner in place. Liners are inexpensive and easy
to replace. A contaminated Detector is a very costly mistake. If unit becomes contaminated
remove liner and clean unit as stated in CHAPTER 20: CLEANING AND MAINTENANCE
before operating.
CAUTION
Care must be exercised when moving the system or when maintenance is performed. The
system must not be inclined more than a 15-degree angle or tip over may occur.
CAUTION
Verify that the Collimator is correctly installed as stated in the Unpacking and Assembly
Instructions. Failure to follow these instructions could result in injury or product malfunction.
CAUTION
This equipment generates radio frequency energy and, if not installed and used in
accordance with the instructions, may cause harmful Electromagnetic Interference (EMI) to
nearby devices. However, there is no guarantee that interference will not occur in a particular
installation.
CAUTION
If any printer other than the model supplied by Capintec is used, the safety of the unit may be
compromised or Electromagnetic Interference (EMI) may be introduced into other devices
located in the same general area as the CAPTUS
®
600 may become susceptible to EMI.
CAUTION
The unit contains lead. The unit should be disposed of in accordance with local and national
regulations.
November 07
SAFETY
1-5
Summary of Contents for CAPTUS 600
Page 2: ......
Page 40: ......
Page 62: ......
Page 77: ...CAPINTEC INC CAPTUS 600 Page 1 November 07 ACCEPTANCE QUALITY ASSURANCE TESTS 6 15...
Page 78: ...CAPINTEC INC CAPTUS 600 Page 2 Page 3 6 16 ACCEPTANCE QUALITY ASSURANCE TESTS November 07...
Page 90: ......
Page 102: ......
Page 112: ...CAPINTEC INC CAPTUS 600 Figure 9 15 9 10 THYROID UPTAKE ONE CAPSULE METHOD November 07...
Page 120: ...CAPINTEC INC CAPTUS 600 Figure 10 11 10 8 THYROID UPTAKE TWO CAPSULES METHOD November 07...
Page 131: ...CAPINTEC INC CAPTUS 600 Figure 11 16 November 07 THYROID UPTAKE STANDARD CAPSULE METHOD 11 11...
Page 132: ......
Page 150: ...CAPINTEC INC CAPTUS 600 Figure 13 11 13 8 WIPE TEST November 07...
Page 152: ......
Page 162: ...CAPINTEC INC CAPTUS 600 Figure 15 5 15 4 LAB TESTS URINE TEST November 07...
Page 167: ...CAPINTEC INC CAPTUS 600 Figure 16 5 November 07 LAB TESTS BLOOD TESTS 16 5...
Page 171: ...CAPINTEC INC CAPTUS 600 Figure 16 9 November 07 LAB TESTS BLOOD TESTS 16 9...
Page 175: ...CAPINTEC INC CAPTUS 600 Figure 16 14 November 07 LAB TESTS BLOOD TESTS 16 13...
Page 176: ......
Page 186: ......
Page 208: ......
Page 214: ......
Page 216: ......
Page 217: ......
Page 218: ......