By appropriately adjusting the “
Delta
” values, it is therefore possible to manage movements and react to sudden events without
waiting for the times introduced by the average.
The device does
NOT
respond to
SHOCK
events and cannot be intended for dynamic event displays. By "sudden event" is meant as
significant displacement related to the position of rest in which the instrument is working.
d.
Reset installation position.
After installation and fixing, a "6 + 1" command can be issued (see
), which resets the channels dedicated
to the inclination.
The command starts a first acquisition which determines the absolute position of the tool and the consequent rotation of the axes
to align the internal references. At the end of the procedure, the zero value and the position of the axes are saved in the memory
so that they can be recovered in case of restart of the device.
This procedure takes about 5 minutes.