10
2B702-068EN*N
Low/Middle-Level Disinfection -1 (Except for USA)
Note that local regulations may require that the transducer be disinfected before sterilization.
Chemical
name/type
Didecyldimethyl-
ammonium chlorides/
Quaternary ammonium
compounds
Didecyldimethyl-
ammonium chlorides
Glyoxal/Glutaral/
Didecyldimethyl-
ammonium chlorides
Trade name
CLEANISEPT
®
WIPES
WIP'ANIOS Excel
Lysoformin
®
3000
Manufacturer
Dr. Schumacher
GmbH
Laboratories
ANIOS
Lysoform
Disinfection Ltd.
Country of origin
DEU
FRA
DEU
Concentration
(Dilution ratio)
Working solution
Working solution
2%
(50)
Time
Wiping only
Wiping only
15 min.
Temperature
Room temperature
Room temperature
Room temperature
Humidity
Normal humidity
Normal humidity
Normal humidity
Model name
Pressure
Normal pressure
Normal pressure
Normal pressure
PSU-25BT/30BT/70BT
PVU-375BT/382BT
Immersible range
OK
OK
OK
Type A
PSU-50BT
PVU-712BT
Type A
OK
OK
OK
PVU-674MV
Type A
–
–
–
PLU-704BT
Type A
OK
OK
OK
PVU-745BTV
PLU-705BTF/705BTH
Type A
OK*
OK
OK
PVU-745BTF/745BTH
Type A
OK*
OK
–
PVU-770ST, PVL-715RS
Type A
–
OK
–
PVU-674BT, PLU-1005BT
Type A
OK
OK
OK
PVU-781VT/781VTE
Type A
–
OK
OK
PLU-1204BT
Type A
OK
OK
–
PVU-681MVL
Type A
–
OK
OK
PVU-350BTP
Type A
–
–
OK
PLU-308BTP
Type A
OK
OK
OK
PLU-1202BT
Type A
OK
OK
OK
OK : Use of the chemical is permitted.
: Use of the chemical is not permitted.
*: Only if the transducer is not used internally.
–
Country of origin :
Any : Any country
FRA : France
DEU : Germany
GBR : United Kingdom
USA : United States
JPN : Japan
AUS : Australia
CAN: Canada