Performing the calibration
After valid calibration data has been obtained by testing calibration
standards, the CCS instrument may be calibrated using the following
procedure:
1. Use the VISCPRO
®
software to log on as a Manager (
Main/Log In
)
in order to access the calibration menu.
2. Click
Configure
from the VISCPRO
®
primary display and select
your CCS instrument.
3. Click
Calibration
. The
Calibration
window will appear.
CCS Calibration window
4. Select the desired calibration
Test Temperature
. If calibration
constants are known, they may be input manually in the appropriate
fields and saved by clicking
Save New Constants
. If test data from
calibration standards is to be used, select the desired date range for