The CCS Data Table window
NOTE
To change the
Tray (Test) Settings
for your instrument(s), you must log in
to the security system as a manager.
4. Edit the values for
Tray (Test) Settings
as necessary. When done,
click OK to save current values and exit the
Tray (Test) Settings
window.
Initial Stepper Motor Service
Prior to performing instrument actions with the CCS-2100 Automatic
Cold-Cranking Simulator, you must first home the Sample Tray mecha-
nism as follows:
1. Make certain that the chiller is connected and operating properly.
Then turn the CCS instrument is
ON
.
2. Load the VISCPRO software by double-clicking the VISCPRO
®
icon
on your Windows
®
desktop (Windows
®
NT
®
users can click
Start/
Programs/VISCPRO/VISCPRO.EXE).
3. Click
Service
from the menu bar and select your CCS instrument
from the list.
4. Select
Stepper Motor Service
from the
Service
options.
5. Click on the
Tube Up
button until the tube has
reached the top of its
travel.
6. Ensure that the arrow on
the Sample Tray is aligned
with the arrow on the
Sample Table (
Table
Forward
/
Table Back
buttons).
7. Click on the
Trolley Left
button until the sampler
has reached the limit of its
travel. If the Sample Tray
has been aligned properly,
the sampler should now be
centered over the #1
sample position.
CCS-2100 with tray in home posi-
CCS sampler mechanism
with tube in home position
(raised)