Alexandrite Long Pulse Laser 755nm-GENTAL MAX PRO
Tables for routine testing and maintenance
Inspection/service
Frequency
Operator
Focus bald clean
After each use
Clinical
departments
External routine cleaning and
disinfection
According to the clinical
protocol
Clinical
departments
Checking whether the cable
and
equipment
surface
damage or not
weekly
Clinical
departments
Checking
the
electrical
connections
weekly
Clinical
departments
Detecting emergency stop
switch
weekly
Clinical
departments
Electrical safety testing
Each year (or in accordance
with the calibration process)
Authorized
personnel of our
company
Test and perform power meter
calibration
Each year, or the system
can't specification,
or An error information
occurred
Authorized
personnel of our
company
6.5
Emergency stop switch test
Detection methods: boot, according to the main interface, when press the stop
button, the system immediately shut off the device; Rotate in the direction shown by the
arrow urgent stop switch button, start equipment
6.6
Replacement of insurance fuse
Steps:
6.8.1
6.8.2
6.8.3
6.8.4
1. Shut down the machine, unplug the power cord from the power socket; The lower
part of the insurance tube position in power. As shown in picture 6.8.1
2. Aim the screwdriver, to dial outside, take out insurance tube module, As shown in
picture 6.8.2
3 .Insurance tube module, there are two fuse. Slot inside the fuse is working state,
behind the fuse box is a backup. Can spare fuse removed, replace the fuse inside the slot.
As shown in picture 6.8.3