Alexandrite Long Pulse Laser 755nm-GENTAL MAX PRO
d. Pulse width selection,
By this page, the pulse width can be selected from 10- 100ms.
e. Frequency selection,
By this page, the frequency can be selected from 0.5-2.0HZ.
f.
By this page, the spot-size can be selected by
6mm(diameter), 8mm, 10mm, 12mm, 15mm, 18mm.
g.
The volume for cold air can be adjusted by three grade.
h.
i.
j.
Switch on or off the cold air cooling for skin.
Press 5 seconds to enter into the engineer mode by password.
To check clients information installed.
k.
Cold air blowing Pulse duration and Pulse interval selection.
The pulse duration and pulse interval can be selected by 0ms, 50ms, 100ms, 500ms,
1000ms. If Pulse duration =1000ms, and Pulse Interval=0ms, it means the cold air is
blowing the skin continually.
l.
The cooling liquid need to be replaced.
m.
Press red circle to simmer up laser, and then press foot-pedal
or finger switch, the laser starts to shoot. Press blue circle, the laser simmer down for
standby station.
Press to shift wavelength from YAG to Alex.
4.
“Alex” page. The alex page shows following,
18 / 32