Alexandrite Long Pulse Laser 755nm-GENTAL MAX PRO
Pow er Failure
Replace the power supply, please contact
the authorized maintenance personnel
Las er ener gy
out put is low
The life of the
light deplet ion
Appropriate to increase energy, can continue
to use
Water quality is not good
High quality water immediately replaced
Laser gener at or
polar ization
Please contact the authorized maintenance
personnel
No
laser
out put ,
but
the
red
aiming beam
nor mal
The life of the
light deplet ion
Replace the pump light source, please
contact
the
authorized
maintenance
personnel
Las er
gener at or
polar ization
please contact the authorized maintenance
personnel
Cont rol panel failu r e
Change control board, please contact the
authorized maintenance personnel
The light path deviation
The light path correction, please contact the
authorized maintenance personnel
No
laser
out put , No
red
aiming
beam out put
laser ar m dam age
Change laser arm, please contact the
authorized maintenance personnel
Cont rol panel failu r e
Change control board, please contact the
authorized maintenance personnel
Water
flow
alarm
Lack of water in the water
tank
The water tank filled with water
With air in the waterway
Confirm the vent plug for free vent plug
Flow sensor dam aged
please contact the authorized maintenance
personnel
Pump does not st art
please contact the authorized maintenance
personnel
There
are
bent
of
waterway system
please contact the authorized maintenance
personnel
Pump damage
please contact the authorized maintenance
personnel
The
water
temperature
alarm
The water temperature is
too high
Wait for cooling, or change of cooling water
Operation
room
temperature is too high
Operation
room ventilation, open air
conditioning (if have)
The water temperature
module damage
please contact the authorized maintenance
personnel
Control panel is damaged
please contact the authorized maintenance
personnel
1).
If the machine is in the process of start or normal operation failure, please stop
running immediately. Unless there is a short of water tank, other fault can only authorized
by the Beijing perfectlaser company maintenance personnel.
2).
If the equipment malfunction, please record the fault state, and contact the
maintenance department of the Beijing perfectlaser company
26 / 32
pum p
pum p