Model 1407 Controls:
PULSE HEIGHT:
5.4
NORMALIZE:
10
POS/NEG:
POS
LINE/OFF/90 Hz:
90 Hz
RISE TIME:
MIN
FALL TIME:
400
µ
s
ATTENUATION:
X10
2.
Connect a 93 or 100 ohm terminator to the 2015A rear panel INPUT.
3.
Connect the 1407 ATTEN OUTPUT to the 2015A front panel INPUT with
RG-62 coax cable.
4.
Connect the 1407 NORMAL OUTPUT to the EXT TRIG input of the scope.
Set the scope triggering to EXT, (+).
Outputs
1.
Connect the AMP OUTPUT to the scope with RG-62 coax cable using a
“tee” connector at the scope. You should observe the near-Gaussian shaped
linear pulse shown in Figure 17.
22
Performance Check
Figure 17 Amp Output Showing a
Near Gaussian Shaped Pulse