+24 V dc
+23.98 to +24.02 V dc
–24 V dc
–23.98 to –24.02 V dc
+12 V dc
+ 11.99 to + 12.01 V dc
–12 D dc
– 11.99 to –12.01 V dc
Current Measurements
Apply power to the 2015A and measure the currents. They should be within the fol-
lowing ranges:
+24 V dc
40 to 0 mA
+12 V dc
150 to 170 mA
– 12 V dc
60 to 80 mA
– 24 V dc
40 to 60 mA
Note:
A greater deviation in currents indicates a faulty unit. Gross errors would
probably be due to faulty or reversed capacitors, shorted or open AMP Out-
put transistors, etc.
Amplifier Operational Checks
Setup
1.
Set controls as follows:
Model 2015A Controls:
COARSE GAIN:
4
FINE GAIN:
10
SHAPING:
2
µ
s (internal)
POLE ZERO:
fully CCW
INPUT POLARITY:
POS
21
Amplifier Operational Checks