Appendix C. Terminal Commands
C-6
GMO – Acquire a new Waveform and return Water Content Value
Usage: GMO\r\n
Response: \r\n[f.ffff] – where f.ffff is an ASCII representation of a floating
point value (4 decimal places)
GVER – Get Operating System Version
Usage: GVER\r\n
Response: \r\n[version] – where version is the current version of the operating
system
GSIG – Get ROM Signature
Usage: GSIG\r\n
Response: \r\nRom Signature:[space][sig] – where sig is the hex
representation of the OS signature
GVAR – Get Variables (Return First Peak and Second Reflection from
previous WC calculation)
Usage: GVAR\r\n
Response: \r\n[f.ffff]
\r\n[f.ffff]
where the first f.ffff returned is an ASCII representation of the value of the first
peak (in met
re
s) and the second f.ffff returned is an ASCII representation of the
value of the second reflection (in met
re
s)
GWA – Acquire and return a Waveform
Usage: GWA\r\n
Response: \r\n[NNNN],[space][f.ffff]
\r\n[NNNN],[space][f.ffff]
\r\n[NNNN],[space][f.ffff]
etc.
where NNNN is the 4-digit width point number and f.ffff is an ASCII
representation of the associated floating point value (4 decimal places)
RSU – Retrieve Current Setup from Flash and Configure Device
Usage:
RSU\r\n
Response:
\r\nSetup has been configured from flash as follows:
\r\n
Vp = [vp value]
\r\n
Ave = [number of averages value]
\r\n
Points = [number of points value]
\r\n
Distance (Cable Length) = [cable length value]
\r\n
Window Length = [window length value]
\r\n
Probe Length = [probe length value]
\r\n
Probe Offset = [probe offset value]
\r\n
Probe Cell Constant = [probe cell constant value]
\r\n
50/60 Hz Rejection = [rejection Hz value]
\r\n
Filter Level = [filter level value]
\r\n Length Apparent Algorithm = [LAA value]
Each line containing a value has an appropriate number of space
characters to ensure that 26 total characters are between the ‘\n’
character and the ‘=’ character.
SSU – Save Current Setup to Flash
Usage:
SSU\r\n
Response: \r\nSetup has been saved to Flash.