Appendix C. Terminal Commands
C-4
SDI – Set Distance
Usage:
SDI[space][value]\r\n
– range of value = [-2.0 – 3822.0]
Response: \r\n>[space]SDI
SWL – Set Window Length
Usage:
SWL[space][value]\r\n
– range of value = [0.0 – 3824.0]
Response: \r\n>[space]SWL
SPL – Set Probe Length
Usage:
SPL[space][value]\r\n
– range of value = [0.0 – 10.0]
Response: \r\n>[space]SPL
SPO – Set Probe Offset
Usage:
SPO[space][value]\r\n
– range of value = [0.0 – 0.50]
Response: \r\n>[space]SPO
SCC – Set Probe Cell Constant
Usage:
SCC[space][value]\r\n
Response: \r\n>[space]SCC
SREJ – Set 50/60 Hz Rejection
Usage:
SREJ[space][value]\r\n
– range of value = [0,50,60]
Response: \r\n>[space]SREJ
SFIL – Set Filter Level
Usage:
SFIL[space][value]\r\n
– range of value = [0 – 10]
Response: \r\n>[space]SFIL
SLAA – Set Length Apparent Algorithm
Usage:
SLAA[space][value]\r\n
– range of value = [0 - 2]
Response: \r\n>[space]SLAA
SMUX – Set SDMX50 channel
Usage:
SMUX[space][value]\r\n
– value = (address[1 to 15]*10) +
channel[1 to 8]
Response: \r\n>[space]SMUX
SDEF – Set Measurement Parameters to Defaults
Usage:
SDEF\r\n
Response:
\r\nSetup has been configured as follows:
\r\n
Vp = [vp value]
\r\n
Ave = [number of averages value]
\r\n
Points = [number of points value]
\r\n
Distance (Cable Length) = [cable length value]
\r\n
Window Length = [window length value]
\r\n
Probe Length = [probe length value]
\r\n
Probe Offset = [probe offset value]
\r\n
Probe Cell Constant = [probe cell constant value]
\r\n
50/60 Hz Rejection = [rejection Hz value]
\r\n
Filter Level = [filter level value]
\r\n Length Apparent Algorithm = [LAA value]
Each line containing a value has an appropriate number of space
characters to ensure that 26 total characters are between the ‘\n’
character and the ‘=’ character.
NOTE