6. Enclosure Humidity
Change the enclosure desiccant packs (model number 4905) when the
enclosure relative humidity exceeds 40%.
Campbell Scientific recommends that all cable entry ports, that do not use a
sealed bulkhead connector, be sealed with plumbers putty and that desiccant
packs be placed inside the enclosure. Spikes in the enclosure humidity are a
result of opening the enclosure door and allowing ambient air inside the
enclosure. The enclosure relative humidity will return its nominal values
(values before the enclosure door was opened) in approximately three to four
hours.
7. Maintenance
The CS210 does not have any user serviceable parts nor does it require any
routine maintenance.
Replace the CS210 probe every three to five years of continuous use. If the
probe fails, replace it with a new one.