TABLE 6-1. Analog Voltage Measurement Accuracy Offsets
Range
(mV)
Differential with
Input Reversal
(µV)
Differential or Single-
Ended without Input
Reversal
(µV)
±5000
±4
±40
±1000
±2
±10
±200
±1
±3
TABLE 6-2. Analog Voltage Measurement Speed
1
f
N1
(Hz)
Multiplexed
2
Measurement
With Input Reversal
Without Input Reversal
Time
(ms)
Rate
(Hz)
Time
(ms)
Rate
(Hz)
30000
1.46
698.49
0.75
1394.05
15000
1.53
667.41
0.78
1332.15
7500
1.66
612.87
0.85
1223.49
3750
1.93
526.78
0.98
1051.89
2000
2.40
422.83
1.22
844.59
1000
3.40
297.18
1.72
593.82
500
5.40
186.39
2.72
372.58
100
21.40
46.81
10.72
93.60
60
34.73
28.82
17.38
57.63
50
41.40
24.18
20.72
48.35
30
68.06
14.70
34.05
29.40
25
81.40
12.29
40.72
24.58
15
134.73
7.42
67.38
14.85
10
201.40
4.97
100.72
9.93
5
401.40
2.49
200.72
4.98
2.5
801.40
1.25
400.72
2.50
1