10.2 Standard output ranges
Soldered jumpers are provided for the coarse setting of the
output ranges and the fine adjustment is accomplished using
the potentiometers “Zero” and “Span”.
Current
[mA]
Soldered jumpers
Voltage
[V]
Soldered jumpers
Channel
1
Channel
2
Channel
1
Channel
2
0 … 20
B120
B220
0 … 10
B120
B122
B123
B220
B222
B223
4 … 20
B121
B221
2 … 10
B121
B122
B123
B221
B222
B223
± 20
–––
–––
± 10
B122
B123
B222
B223
10.3 Specific user output ranges
Units that have been configured for a specific user output
range cannot be subsequently reconfigured.
B21
2
B21
3
B21
4
B21
5
B21
6
B21
7
B207
B208
B209
B210
B211
B201
B202
B203
B204
B205
B221
B220
B122
B123
B113
B114
B115
B116
B117
B107
B108
B109
B110
B111
B101
B102
B103
B104
B105
B206
B106
Span 1
Zero 1
Span 2
Zero 2
B121
B120
B223
B222
B112
Fig. 6. Position of the soldered jumpers B…, potentiometers
“Zero” and “Span”.
11. Mounting
The isolating amplifier SIRAX TV 808 is plugged onto a
backplane BP 902.
When deciding where to install the transmitter
(measuring location), take care that the
limits
of
the operating temperature
are kept:
– 25 and + 40 °C for standard instruments!
11.1 Plugging the module into the backplane
Before inserting the SIRAX TV 808 into the
backplane, ensure that, …
… the backplane wiring is in strict accordance
with the wiring diagram of the module
… the backplane is coded correctly ac-
cording to the section entitled “Mecha-
nical coding of the backplane”. Read the
instructions for the back-plane.
… The red coding insert has been remo-
ved from the backplane for SIRAX plug-in
modules with a power supply of 24…60 V
DC/AC and that the power supply is correct
for the module.
1. Clip the module base onto the top-hat rail.
2. If the backplane is mounted vertically, turn the quick release
screws on the module to a vertical position, respectively if
it is mounted horizontally, turn the screws to a horizontal
position.
3. Press the quick release screws inwards with the screw-
driver until there is an audible click.
SIRAX
3.
GOSSEN
METRA
WA
TT
C
AM
ILLE BAUER
SIRAX
1.
2.
Fig. 7. Plug the module into the base.
12. Commissioning
Switch on the measuring inputs and the power supply. The
green LED lights continuously after switching on.
The power supply unit must be capable of supp-
lying a brief current surge when switching on. The
isolating amplifier presents a low impedance at
the instant of switching which requires a current
I
start
of …
… I
start
≥
160 mA for the version with a power supply
range of 24 – 60 V DC/AC
or
… I
start
≥
35 mA for the version with a power supply
range of 85 – 230 V DC/AC