Cambridge Sensotec Rapidox 1100E Oxygen Analyser Instruction Manual
Version 3.8
36
Last printed 29/08/2012 16:01:00
Figure 19: Rapidox Diagnostics Page
1.
Analogue
:
The live analogue values displayed in this pane are voltage
measurements coming directly from the computer chip inside the analyser.
They are for information purposes only but are a useful diagnostics tool for
Sensotec staff, as unusual values can indicate a particular fault. To help
Sensotec diagnose a fault please capture this window as shown above using
the screen capture function described in section 7.9 above and e-mail it to
[email protected]
2.
Sensor State: These boxes should all indicate “OK” on a healthy analyser.
The possible messages are “OK” “High” “Low” or “Fault”. The High and
Low messages indicate that the sensor is out off normal range. The fault
message will only display if the sensor is disconnected or has failed in some
way.
3.
Digital: The digital signals coming directly from the computer chip inside
the analyser can be monitored here. To test the keypad is working correctly
try pushing the buttons and watching the panel. A correctly functioning
key will turn the box red on the screen and the digital value for that
particular key will show in the data box. If the analyser has an external
thermocouple correctly fitted (2100 and 3100 series only) then this will
display a signal in the “Temperature Sensor” box.
4.
Log Diagnostic Data: Checking this box will allow the software to log the
diagnostic data which can then be e-mailed to Sensotec for analysis. Do not
switch to a different tab while logging diagnostic data – the data is only
Summary of Contents for SENSOTEC Rapidox 1100E