Cambridge Sensotec Rapidox 1100E Oxygen Analyser Instruction Manual
Version 3.8
20
Last printed 29/08/2012 16:01:00
The user can choose whether to use ppm or percent notation to edit the O
2
text
boxes by selecting the UNITS menu and clicking on ‘Percent’ or ‘ppm’. To access
a field, either click inside it or use the TAB key to scroll through them. The field
parameters are as follows:
7.5.1
Config 1 Screen
1)
Oxygen Range: Not available for the Rapidox 1100E
2)
Oxygen Units: See Section 6.9 for a full description. Use the drop down
box to select AUTO%, PPM, PRESSURE or N2 BALANCE. The default
setting is AUTO%.
3)
Temperature Units: Not available for the Rapidox 1100E.
4)
Pressure Units: See sections 6.9 for a full description. Use the drop down
menu to select MBAR, BAR, PSI, TORR or KPa as the unit for pressure
measurement. The default setting is MBAR.
5)
Pressure Correction Mode: See section 6.12 for a full description. Use the
drop down menu to select NONE or AUTOMATIC. The default setting is
AUTOMATIC.
6)
Operating Mode: Not available for the Rapidox 1100E
7)
Auxiliary Output Parameter: This is fixed to PRESSURE on the Rapidox
1100E and is therefore not adjustable in the software.
8)
LCD Refresh interval: Use the drop down menu to select a refresh rate for
the LCD (min 0.1, max 1.5 sec). The default is 0.3 seconds.
9)
LCD Second Line Option: See section 6.10 for a full description. Use the
drop down menu to select Pressure, Date/Time or None.
10)
Pump Flow Rate: See section 6.4 for a full description. Use the drop down
menu to select a pump flow value from 10 to 100% or OFF.
11)
Password option: See section 6.5 for a full description. The password is
factory disabled and set to ‘0000’. Use the drop down menu to enable or
disable the password feature and the password box ‘****’ can be used to
enter a new password. Note that you will be prompted to enter the original
password if you wish to make changes to the password or its status.