450866-001 Rev. B                                      caltestelectronics.com                                                            8

Vertical Scale on Oscilloscope

The actual vertical scale of the oscilloscope is equal to the attenuation 

factor multiplied by the range of vertical scale selected on the 

oscilloscope. For example, with the probe on factor 10x and the 

oscilloscope on 0.5 V/div, the real vertical scale is 10 x 0.5 = 5 V/div. 

With the probe on 100x, the real vertical scale is 100 x 0.5 = 50 V/

div. These values apply when the oscilloscope is set to the typical 1 

MΩ impedance input. When the scope is set to 50 Ω input, the actual 

vertical scale will be doubled: 10 V/div for the 10x setting and 100 V/div 

for the 100x setting.  See the chart below.

Vertical Scale on Oscilloscope

Scope Input 

Impedance

Probe 

Attenuation 

Setting

Actual 

Attenuation 

Setting

Vertical Scale 

Reading on the 

Oscilloscope

Actual Vertical 

Scale of the 

Oscilloscope

1 MΩ

10x

10x

0.5 V/div

5 V/div

1 MΩ

100x

100x

0.5 V/div

50 V/div

50 Ω

10x

20x

0.5 V/div

10 V/div

50 Ω

100x

200x

0.5 V/div

100 V/div

  

 

Table 1  Oscilloscope Readings

Summary of Contents for CT4192

Page 1: ...CT4192 User Manual...

Page 2: ...ne Do not work alone when working with high voltages Inspect the Probe Inspect the probe and accessories for cracks and frayed or broken leads before each use If defects or damages are noted DO NOT US...

Page 3: ...r Not for Critical Applications This probe is not authorized for use in contact with the human body or for use as a component in a life support device or system Do Not Substitute Parts Do not install...

Page 4: ...ainst any unintended operation Compliance Statements Disposal of Old Electrical Electronic Equipment Applicable in the European Union and other European countries with separate collection systems This...

Page 5: ...3dB z z Up to 700 V DC AC peak common mode z z Selectable attenuation settings of 10x 100x z z Power indicator z z High accuracy 1 z z Powered by 4 AA batteries included z z Power adapter CT3723 optio...

Page 6: ...erview CT 4192 Description CT4192 1 M Hook probes black and red Attenuation Function Switch Output BNC Connector Power On Off Switch Power Indicator Red Offset Adjust External Power Connection Figure...

Page 7: ...teries being sure to match polarity with the printed indicators inside the compartment Then Slide the cover back and replace the rubber boot See Figure 2 AA AA AA AA Figure 2 Battery Connection Altern...

Page 8: ...V switch the attenuation ratio to 10x in order to get higher resolution and less noise ratio For voltages up to 700 V set the attenuation ratio to 100x 4 Plug the hook probes onto the test leads and c...

Page 9: ...5 50 V div These values apply when the oscilloscope is set to the typical 1 M impedance input When the scope is set to 50 input the actual vertical scale will be doubled 10 V div for the 10x setting...

Page 10: ...Default Setup and Auto on the oscilloscope 5 Press the Channel 1 button then press the probe softkey and set the attenuation to match that of the probe 6 Set the oscilloscope to DC coupled mode and t...

Page 11: ...ntial Input Voltage DC AC peak 70 V 10x 700 V 100x Maximum Common Mode Input Voltage DC AC peak 70 V 10x 700 V 100x Absolute Max Rated Input Voltage each side to ground 1000 Vrms Input Resistance Capa...

Page 12: ...de Operating 3 000 m Nonoperating 15 300 m Safety Specifications IEC 61010 031 CAT II Mechanical Characteristics Weight 400 g with probe and rubber boot Dimensions 170 x 63 x 21 mm BNC Cable Length 95...

Page 13: ...Derating Curve The derating curve of the absolute maximum input voltage in common mode is shown as follows 1 000 10 000 100 10 1 10 100 Frequency MHz Volgage Vrms CT4192 Maximum Input Voltage Deratin...

Page 14: ...persons will not be covered by this warranty Cal Test Electronics reserves the right to discontinue models change specifications price or design of this device at any time without notice and without...

Page 15: ...450866 001 Rev B caltestelectronics com 14 Notes...

Page 16: ...2017 Cal Test Electronics 450866 001 Rev B...