User and Programming Manual - Rev H
California Instruments
May 2000
P Series
4
Current:
6.0 A rms, 17.8 A peak (low range, model 801P)
3.0A rms, 8.9 A peak (high range, model 801P)
7.4 A rms, 22.2 A peak (low range, model 1001P)
3.7 A rms, 11.1 A peak (high range, model 1001P)
9.2 A rms, 27.6 A peak (low range, model 1251P)
4.6 A rms, 13.8 A peak (high range, model 1251P)
Current Limit:
801P
1001P
1251P
135V range:
0.0 to
6.0
7.4
9.2
270V range:
0.0 to
3.0
3.7
4.6
Accuracy:
Programmed value +5% of maximum current
Frequency Range:
Range
Resolution
16.0 - 99.9 Hz
0.1 Hz
100 - 500 Hz
1 Hz
Frequency Accuracy:
±
0.02% of program value
DC Offset Voltage:
Less than
±
25 mV with linear load.
Output Noise:
<0.2 volts RMS on 135 range, <0.5 volts RMS on 270 range.
2.1.3 Measurements
Current (TRMS)
Resolution:
0.1 amp
Accuracy:
±
0.2 amp
Voltage
Accessible only through RS232
0 - 250V
251 to 270V
Resolution
1 Volts
2 volts
Accuracy:
±
4 Volts
±
6 volts