SettingS
date & Time
Change the date and time for data logging purposes. The clock is reset when
the power is cycled unless a battery is fitted.
Output processing
averaging
period
hold Mode
hold period
Set the time, in seconds, over which the measured temperature is averaged.
Note: averaging prevents the sensor from following rapid temperature
changes. Minimum: 0 (no averaging). Maximum: 60.
peak
The sensor holds the maximum temperature steady for the Hold Period.
After this, the sensor responds normally. If the sensor detects a higher
temperature, it holds this temperature steady for the Hold Period.
Valley
Similar to Peak Hold mode except that the sensor outputs the minimum
detected temperature steady for the Hold Period.
Off
Disables hold processing.
Set the time, in seconds, for the sensor to hold the temperature as above.
Minimum: 0 (no hold processing). Maximum: 1200.
data logging
sample
period
number of
samples
enable
scheduled
start
date and
Time
The time, in seconds, between samples. Minimum: 1. Maximum: 86,400
(1 day).
The number of samples to collect before logging stops. Minimum: 0
(continuous logging). Maximum: 86,400 (1 day of data if Sample Period = 1
second).
The sensor begins logging at the Date and Time specified. Logging can also
be started and stopped manually.
The date and time for scheduled logging to start.