Calex ExTempMini Series Quick Start Manual Download Page 2

INTRODUCTION

ExTempMini intrinsically safe non-contact infrared temperature sensors measure the
temperature of an area of the surface of a solid or liquid, and transmit this as a linear 4-20 mA
output.

Temperature ranges from -20°C to 1000°C are available. All models have an adjustable
emissivity setting, and may be used to measure a wide variety of target materials, including
food, paper, textiles, plastics, leather, tobacco, pharmaceuticals, chemicals, rubber, coal,
asphalt and paint.

A choice of optics is available to measure small or large targets at short or long distances.

The optional LCT Loop Configuration Tools (USB and RS-485 adapters) and free software allow
the ExTempMini to be connected to a PC, PLC or SCADA system for temperature indication,
sensor configuration and data acquisition.

These sensors are designed primarily for use in hazardous areas in conjunction with a
suitable Intrinsically Safe isolator. All models have been certified Intrinsically Safe for use in
gas hazardous areas by Certification Management Ltd. They comply with the European ATEX
Directive 2014/34/EU and are covered by certificates for IECEx (international) and UKCA
(United Kingdom). The ExTempMini Series is also approved for use in Japan (JNIOSH-
TR-46-1:2020 (General requirements); JNIOSH-TR-46-6:2015 (Intrinsic Safety “i”)).

IMPORTANT INFORMATION FOR USE:

Voltage

MUST

be supplied by a suitably rated safety isolator.

For re-configuration of the sensor, the LCT

MUST

be connected in the safe area, behind the

protection of a safety isolator.

The sensing head

MUST NOT

be opened. Care should be taken to avoid inadvertently

loosening the cable gland when tightening locknuts.

Do not attempt to repair a faulty unit. Contact the vendor to arrange a return.

CONFORMANCE TO REQUIREMENTS

The sensing head has passed a high voltage withstand test up to 700 VDC.

SPECIFIC CONDITIONS OF USE

The main enclosure of this equipment is not capable of withstanding the 500 VAC / 700 VDC
insulation test required by IEC 60079-11:2011 Clause 6.3.13. This shall be taken into account
when installing the equipment. Power to the equipment must be supplied by an intrinsically
safe isolator. The equipment is not compatible with intrinsically safe Zener barriers.

CONDITIONS FOR SAFE USE (JAPAN)

i.

The main enclosure does not meet a 500 Vrms dielectric strength test. This shall
be taken into account when installing the equipment.

ii.

The supply cable used shall have a temperature rating > +70 °C.

SAFETY PARAMETERS:

The device must not be used outside of the ambient temperature range (Ta), or subjected to
voltages, current or power greater than those listed below, in order to ensure safe operation
of the device:

Ui

=

28 V

Ta

=

-20°C to +70°C (Main Enclosure)
-20°C to +180°C (Sensing Head)
(see Hazardous Area Classification)

Ii

=

93 mA

Ci

=

5.17 nF

Pi

=

0.651 W

Li

=

1.99 µH

2

Summary of Contents for ExTempMini Series

Page 1: ...ExTempMini Series Intrinsically Safe Miniature Infrared Temperature Sensor with Separate Electronics Module Operator s Guide...

Page 2: ...ty i IMPORTANT INFORMATION FOR USE Voltage MUST be supplied by a suitably rated safety isolator For re configuration of the sensor the LCT MUST be connected in the safe area behind the protection of a...

Page 3: ...rsist for a short period NO Group Definition Supported Gas IIA e g Propane IIB e g Ethylene IIC e g Hydrogen Dust IIIA Combustible flyings NO IIIB Non conductive dusts NO IIIC Conductive dusts NO Tabl...

Page 4: ...ge may be rescaled between limits 20 C and 1000 C via optional adapters and software Sensing Head Cable Length m e g 5 5 m 10 10 m 25 25 m maximum length The sensing head and electronics module are ca...

Page 5: ...ntal Rating IP65 NEMA 4 IP65 NEMA 4 Ambient Operating Temperature Range 0 C to 70 C 0 C to 180 C Ambient Pressure Range 80 kPa 0 8 bar to 110 kPa 1 1 bar 80 kPa 0 8 bar to 110 kPa 1 1 bar Relative Hum...

Page 6: ...8 V Maximum Input Current Ii 93 mA Maximum Input Power Pi 0 651 W Maximum Internal Capacitance Ci 5 17 nF Maximum Internal Inductance Li 1 99 H ATEX Certificate Number CML 22ATEX2007X IECEx Certificat...

Page 7: ...ended sensing head cable 25 m max factory fitted contact Calex for information about extending the cable Output cable 25 m max factory fitted cable alternatively customer may supply and fit output cab...

Page 8: ...ontaminate the lens and cause errors in temperature measurement In these types of environment the air purge collar should be used to help keep the lens clean Electrical Interference To minimise electr...

Page 9: ...way from the lens It must be screwed on fully There are two models of air purge collar to suit the different optics see below Air flows into the hose barb fitting and out of the front aperture Air flo...

Page 10: ...adapter PC with configuration software Load measurement instrumentation 220 Vs Do not open PWR PWR Sensing head Ta 20 C to 180 C Certified Intrinsically Safe isolator e g connections shown are for MT...

Page 11: ...MODBUS The optional LCT 485 Network Interface not shown allows an ExTemp sensor to be connected to an RS 485 Modbus RTU network for temperature measurement configuration and data acquisition Each LCT...

Page 12: ...ing can be applied by setting Hold Mode to Peak or Valley and setting the hold period This is useful if the temperature reading is interrupted by gaps between moving objects or by an obstruction Defau...

Page 13: ...for further advice LENS CLEANING Keep the lens clean and dry at all times Any foreign matter on the lens would affect measurement accuracy Blow off loose particles if not using the air purge accessory...

Page 14: ...safe side of a suitable certified Intrinsically Safe isolator Do not attempt to repair a faulty unit Contact the vendor to arrange a return ELECTRICAL CONNECTION Connect the device to the 4 to 20 mA...

Page 15: ...ARE See Configurable Parameters and the Configuration Software diagram opposite If the Settings icon is greyed out the software is locked Unlock the software to allow access to the Settings menu The d...

Page 16: ...twork interface Select COM port then select LCT 485 on the next screen Emissivity Compensation Temperature List Temperature Display Output Range Output Processing Default Password 1234 Connection List...

Reviews: