CALIBRATION
TW SERIES
3-22
Calculate the measurement system frequency compensation values:
VALUE1
=
RDGN1
/
MEAS1
VALUE2
=
RDGN2
/
MEAS2
VALUE3
=
RDGN3
/
MEAS3
VALUE4
=
RDGN4
/
MEAS4
VALUE5
=
RDGN5
/
MEAS5
Update the TW’s calibration data with the following GPIB command:
CAL
{1,2,3}
:MEAS:REM:HIR:VOLT:FREQCAL
VALUE1 VALUE2 VALUE3 VALUE4 VALUE5
3.15 EXTERNAL PHASE REFERENCE CALIBRATION
Connect the counter timer channel 1 to the TW’s A phase through a 10:1 resistor
divider. Connect the counter timer channel 2 to the TW’s clock/lock/sync BNC. Set the
counter time up to measure phase angle.
Set up the TW for the external refout, low range, local sense, 40 Hz, 120 V and close
the output relay by sending the following GPIB commands:
SYST:EXT:CLOCK REFOUT
SOUR:VOLT:RANGE LOW
SOUR:SENSE LOCAL
SOUR:FREQ 40.0
SOUR:VOLT 120.00
OUTPUT ON
Wait for the phase angle measurement to stabilize on the counter timer. Record the
phase angle as
RDGN1
.
Set the TW frequency to 80 Hz:
SOUR:FREQ 80.0
Wait for the phase angle measurement to stabilize on the counter timer. Record the
phase angle as
RDGN2
.
Set the TW frequency to 160 Hz:
SOUR:FREQ 160.0
Wait for the phase angle measurement to stabilize on the counter timer. Record the
phase angle as
RDGN3
.
Set the TW frequency to 160 Hz:
SOUR:FREQ 160.0
Summary of Contents for TrueWave TW5250
Page 2: ......
Page 4: ...ii This page intentionally left blank...
Page 10: ...TABLE OF CONTENTS TW SERIES viii This page intentionally left blank...
Page 36: ...MAINTENANCE AND TROUBLESHOOTING TW SERIES 2 16 This page intentionally left blank...
Page 80: ...CALIBRATION TW SERIES 3 44 This page intentionally left blank...