background image

Operation Manual for

PA1 Power Amplifier

www.bartingtondefenceandspace.com

www.bartingtondefenceandspace.com

Summary of Contents for PA1

Page 1: ...Operation Manual for PA1 Power Amplifier www bartingtondefenceandspace com www bartingtondefenceandspace com...

Page 2: ...4 4 1 Pin Outs 8 4 5 PA1 Power Amplifier Power 9 4 6 Amplifier O P Circuit Breakers 9 4 7 Cooling 10 4 8 Static Field Compensation 10 4 8 1 Method 11 5 Coil Calibration and Alignment Adjustment 11 5 1...

Page 3: ...brackets refers to that label Bartington Defence Space cannot advise on the integration of this equipment with any third party products 1 1 Symbols Glossary The following symbols used within this manu...

Page 4: ...ng Should this occurs the power will to the PA1 will be cut In such event it is advised to allow the unit to cool down before restarting it To prevent the thermal cut out from activating the equipment...

Page 5: ...per front panel 2 Current control input takes input from signal generator 3 Calibration controls behind transparent plastic cover 4 Current monitor output 5 Static field offset controls 6 Fastening sc...

Page 6: ...OM3225 1 3 1 2 PA1 Power Amplifier Rear View Key 11 Blank ventilation panel 12 Fastening screws 13 Coil drive output to the HC1 or HC2 Helmholtz Coils 14 Lower rear panel 15 Circuit breaker 16 Mains p...

Page 7: ...ure should not exceed the maximum specified in the data sheet DS2613 4 2 Connecting to the HCS1 System The PA1 Power Amplifier connects to the HC1 or HC2 Helmholtz Coils and the CU1 Control Unit The H...

Page 8: ...input 2 The current monitor output 4 located on the upper front panel of the PA1 Power Amplifier is also connected to the CU1 Note A diagram showing the interconnection of components of the HCS1 Syst...

Page 9: ...e time of ordering A 16A circuit breaker is situated on the rear panel 15 which is suitable for the supply stated in the product brochure Contact Bartington Defence Space for advice if a lower voltage...

Page 10: ...9 should be inspected regularly and cleaned or replaced when necessary See Air Intake Filter Cleaning and Replacement 4 8 Static Field Compensation The static field is offset by introducing a DC offs...

Page 11: ...uld be checked regularly and will certainly need setting on installation or should the coil or metalwork in the vicinity of the coil be moved or should other sources of magnetic fields be introduced i...

Page 12: ...e use of a true RMS reading voltmeters will enable the user to measure only the AC component of a signal For HC1 a 1Vrms input signal on the PA1 will correspond to a 100 Trms field The output of the m...

Page 13: ...then adjust the trim control until the same level is observed when the magnetometer is in the normal position and rotated position Y axis alignment trim in X axis Feed a signal from the signal genera...

Page 14: ...nnector can then be disconnected if required Fan filter maintenance must be performed by suitably trained personnel only Remove the power connector from the lower rear panel before removing the centre...

Page 15: ...e any other connection to these ports 8 Troubleshooting Care and Maintenance In the event of any apparent malfunction email service bartingtondefenceandspace com or telephone the Bartington Defence Sp...

Page 16: ...10 1 Waste Electrical and Electronic Equipment WEEE Regulations This product complies fully with Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment RoHS and...

Page 17: ...nts of the HCS1 Helmholtz Coil System PXI CHASSIS CU1 PA1 DAQ DAQ Connector 0 Connector 1 Current Control Output Aux Input Current Monitor Input Device Test Input Current Control Input Current Monitor...

Page 18: ...specific versions for most countries 3 If an un terminated IEC connector lead is supplied with the equipment the wires must be wired accordingly BROWN LIVE BLUE NEUTRAL GREEN YELLOW EARTH 4 It is ess...

Page 19: ...W www bartingtondefenceandspace com www bartingtondefenceandspace com OM3225 1 The copyright of this document is the property of Bartington Defence and Space Ltd Bartington is a registered trade mark...