8 - ENG PN MN239 F | User Manual
NewLife Intensity 10
CAUTION: Normally, you should not need to adjust
the flowmeter on your unit. If you turn the flowme-
ter adjustment knob clockwise, you will decrease
and can shut off the flow of oxygen from your unit.
For your convenience, the flowmeter is marked in
½ LPM increments. For units equipped with the 2
LPM flowmeter option, the flowmeter is marked in
1/8 LPM increments for flow settings up to 2 LPM.
Note: The Manufacturer does not recommend the
sterilization of this equipment.
Note: If the unit has not been used for an extended
period of time, it needs to operate for several minutes
before power failure alarm can become activated.
Note: The concentrator releases warm air out the
bottom of the unit which can permanently discolor
temperature sensitive flooring surfaces such as vinyl.
The concentrator should not be used over flooring that
is sensitive to heat staining. The Manufacturer is not
responsible for flooring that becomes discolored.
Note: The NewLife Intensity Oxygen Concentrator
must be operated for at least five minutes at 2 LPM
before using the unit.
The NewLife Intensity is appropriate for usage by two
users, provided the combined flow is a minimum of 2
LPM and does not exceed the maximum capacity of
the concentrator.
Note: To prevent a void warranty, follow all manufac-
turers’ instructions.
Note: Do not attempt any maintenance other than the
possible solutions listed within the manual.
Note: Portable and mobile radio frequency (RF) com-
munications equipment can effect medical electrical
equipment.
Note: To Equipment Provider: The following oxygen
administration accessories are recommended for use
with the NewLife:
• Nasal Cannula: CAIRE Part Number CU002-1
• Humidifier Adaptor Tubing: CAIRE Part Number
20843882
• Humidifier Bottle: CAIRE Part Number HU003-1
• Firebreak: CAIRE Part Number 20629671
A firebreak is required for use with any cannula.
• CAIRE offers a firebreak intended to be used in con-
junction with the oxygen concentrator. The firebreak is
a thermal fuse to stop the flow of gas in the event that
the downstream cannula or oxygen tubing is ignited
and burns to the firebreak. It is placed in-line with the
nasal cannula or oxygen tubing between the patient
and the oxygen outlet of the NewLife. For proper use
of the firebreak, always refer to the manufacturer’s
instructions (included with each firebreak kit).
• Additional recommended accessories information is
available online at www.caireinc.com.
WARNING: KEEP OUT OF THE REACH
OF CHILDREN UNTIL INSTALLED.
WARNING: THIS PRODUCT CAN EXPOSE YOU
TO CHEMICALS INCLUDING NICKEL, WHICH
IS KNOWN TO THE STATE OF CALIFORNIA TO
CAUSE CANCER. FOR MORE INFORMATION, GO
TO WWW.P65WARNINGS.CA.GOV.