PN MN239 F | User Manual ENG - 23
NewLife Intensity 10
Guidance and Manufacturer’s Declaration± Electromagnetic Immunity
The NewLife is intended for use in the electromagnetic environment specified below. The customer or the user of
the NewLife should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment —
guidance
Electromagnetic
environment – guid-
ance
IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power supply
lines
±1 kV for input/output
lines
±2 kV for power supply
lines
N/A
Mains power quality should be that of a
typical commercial or hospital
environment.
Surge
IEC 61000-4-5
± 2kV common mode
on AC lines
± 1kV differential on
AC lines
±- 2kV common mode
on outdoor I/O lines
N/A
± 1kV differential on
AC lines
N/A
Mains power quality should be that of a
typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines IEC
61000-4-11
0% U
T
for 0.5 cycles
(0°, 45°, 90°, 135°,
180°, 225°, 270°, 315°)
0% U
T
for 1 cycle (0°)
70% U
T
(30% dip in U
T
)
for 25/30 cycles (0°)
0% U
T
for 250/300
cycles (0°)
0% U
T
for 0.5 cycles
(0°, 45°, 90°, 135°,
180°, 225°, 270°, 315°)
0% U
T
for 1 cycle (0°)
70% U
T
(30% dip in U
T
)
for 25/30 cycles (0°)
0% U
T
for 250/300
cycles (0°)
Mains power quality should be that of a
typical commercial or hospital environ-
ment. If the user of the NewLife Intensity
10 requires continued operation during
power mains interruptions, it is recom-
mended that the NewLife is powered
from an uninterruptible power supply
(UPS) or a battery.
Power frequency
(50/60 Hz) magnetic
field IEC 61000-4-8
30 A / m
30 A / m
Power frequency magnetic fields should
be at levels characteristic of a typical lo-
cation in a typical commercial or hospital
environment.
NOTE U
T
is the A.C. mains voltage prior to application of the test level.