PN MN239 F | User Manual    ENG - 17

NewLife Intensity 10

Cleaning, Care, and Proper Maintenance

WARNING: CLEAN THE CABINET, 

CONTROL PANEL, AND POWER CORD 

ONLY WITH A MILD HOUSEHOLD 

CLEANER APPLIED WITH A DAMP (NOT 

WET) CLOTH OR SPONGE, AND THEN 

WIPE ALL SURFACES DRY. DO NOT 

ALLOW ANY LIQUID TO GET INSIDE 

THE CONCENTRATOR. PAY SPECIAL 

ATTENTION TO THE OXYGEN OUTLET 

FOR THE CANNULA CONNECTION TO 

MAKE SURE IT REMAINS FREE OF 

DUST, WATER, AND PARTICLES.

Cabinet

Turn OFF the unit and disconnect from power 

before any cleaning or disinfection. DO NOT spray 

the outer case directly. Use a damp (not wet) cloth 

or sponge. Spray the cloth or sponge with a mild 

detergent solution to clean the cabinet. Proceed as 

directed by the cleaner manufacturer. Device cabi-

net should be cleaned at minimum between users.

WARNING: ELECTRICAL SHOCK 

HAZARD. TURN OFF THE UNIT AND 

DISCONNECT THE POWER CORD 

FROM THE ELECTRIC OUTLET BEFORE 

YOU CLEAN THE UNIT TO PREVENT 

ACCIDENTAL ELECTRICAL SHOCK AND 

BURN HAZARD. ONLY YOUR EQUIP-

MENT PROVIDER OR A QUALIFIED SER-

VICE TECHNICIAN SHOULD REMOVE 

THE COVERS OR SERVICE THE UNIT.

WARNING: CARE SHOULD BE TAKEN TO 

PREVENT THE OXYGEN CONCENTRATOR 

FROM GETTING WET OR ALLOWING FLUIDS TO 

ENTER THE UNIT. THIS CAN CAUSE THE UNIT TO 

MALFUNCTION OR SHUT DOWN, AND CAUSE AN 

INCREASED RISK FOR ELECTRICAL SHOCK OR 

BURNS.
WARNING: DO NOT USE OIL, GREASE, OR 

PETROLEUM-BASED OR OTHER FLAMMABLE 

PRODUCTS WITH THE OXYGEN-CARRYING AC-

CESSORIES OR THE OXYGEN CONCENTRATOR. 

OXYGEN ACCELERATES THE COMBUSTION OF 

FLAMMABLE SUBSTANCES.

NewLife Intensity 10:

Concentrator casing .......... Valtra/ABS/Polystyrene

Mains cable ......................................................PVC

Dust filter ..................................................Polyester

ON/OFF switch ................................ Thermoplastic

Casters ............................................................Nylon

Flow adjustment .......................ABS/Polycarbonate

Gas outlet ...............................Chrome Plated Brass

Printed labels ..................................................Lexan

Materials in Direct or Indirect Contact with Operator

Summary of Contents for NewLife Intensity 10

Page 1: ...NewLife Intensity 10 User Manual English...

Page 2: ...Type BF applied part degree of protection against electric shock Reg 5333 Warning Reg W001 Council Directive 93 42 EEC concerning medical devices Authorized representative in the European Community I...

Page 3: ...home when traveling or while participating in daily activities away from home Oxygen is a gas that makes up 21 of the room air we breathe Our bodies depend on a steady supply to function properly You...

Page 4: ...n operate at any location where AC power source is available Oxygen concentrator use requires a physician s prescription and is not intended for life support use Although oxygen therapy can be prescri...

Page 5: ...perating Instructions Explains procedures to operate the unit H Air Intake Gross Particle Filter Prevents dust and other airborne particles from entering the unit I Power Cord Allows connection of uni...

Page 6: ...RDOUS THIS DEVICE SHOULD ONLY BE USED WHEN PRESCRIBED BY A PHYSICIAN WARNING NOT FOR USE IN THE PRESENCE OF FLAMMABLE ANESTHETICS WARNING AS WITH ANY ELECTRICALLY POWERED DEVICE THE USER MAY EXPERIENC...

Page 7: ...ntal by order of a physician or other licensed health care provider CAUTION Do not position the unit so that it is difficult to access the power cord CAUTION The concentrator should be located as to a...

Page 8: ...m capacity of the concentrator Note To prevent a void warranty follow all manufac turers instructions Note Do not attempt any maintenance other than the possible solutions listed within the manual Not...

Page 9: ...Pressure 20 psig 138 kPA Operating Environment 41 F to 104 F 5 C to 40 C 15 90 humidity Altitude 1250 to 10 000 ft 381 to 3048 m tested to 700 1060 hPa Storage Environment 25 C to 70 C 13 F to 158 F...

Page 10: ...if required nasal cannula face mask catheter and or extension tubing to the oxygen outlet 5 Completely unwrap the power cord 6 Insert power cord into the electrical outlet 7 Locate the power switch on...

Page 11: ...BE USED OXYGEN AC CELERATES THE COMBUSTION OF FLAMMABLE SUBSTANCES WARNING THE USE OF SOME OXYGEN ADMIN ISTRATION ACCESSORIES NOT SPECIFIED FOR USE WITH THIS OXYGEN CONCENTRATOR MAY IMPAIR ITS PERFOR...

Page 12: ...If your physician did not prescribe a humidi fier connect the oxygen tubing directly to the unit s oxygen outlet A separate outlet fitting is supplied for this type of connection Operating With Humidi...

Page 13: ...pter or humidifier Note To Equipment Provider The following oxygen administration accessories are recommended for use with the NewLife Oxygen Concentrator Nasal Cannula with 7 feet 2 1 m of tubing 6 L...

Page 14: ...ermal switch located within the compressor trips shutting down the compressor Oxygen Monitor In the event the oxygen monitor detects an oxygen concentration below 82 the low oxygen concentration warni...

Page 15: ...cation as prescribed by your physician Note Use only the amount of medication and frequen cy of treatment that your physician prescribed 3 Remove or unscrew the medication cup on the nebulizer and pl...

Page 16: ...er parts 1 After each treatment separate the nebulizer and the T piece or mouthpiece assembly 2 Remove or unscrew the nebulizer cup and rinse each component thoroughly in warm water 3 Once a day clean...

Page 17: ...ICAL SHOCK HAZARD TURN OFF THE UNIT AND DISCONNECT THE POWER CORD FROM THE ELECTRIC OUTLET BEFORE YOU CLEAN THE UNIT TO PREVENT ACCIDENTAL ELECTRICAL SHOCK AND BURN HAZARD ONLY YOUR EQUIP MENT PROVIDE...

Page 18: ...least one time each week wash the air intake gross particle filter which is located in the back of the unit Your Equipment Provider may advise you to clean it more often depending upon your operating...

Page 19: ...d intermittent audible alarm Power Failure Ensure device is plugged into a known working outlet Ensure breaker switch is pushed in If issue persists contact equipment provider for service Troubleshoot...

Page 20: ...e unit is positioned away from curtains or drapes hot air registers heaters and fireplac es Be certain to place the unit so all sides are at least 12 inches 30 5 cm away from a wall or other obstructi...

Page 21: ...ard Accessories Nasal Cannula with 7 feet 2 1 m of tubing 6 LPM max CU002 1 Oxygen Outlet Adapter F0025 1 Face Mask with 7 feet 2 1 m of tubing 10 LPM max MS013 1 Humidifier Adapter Extension HU002 1...

Page 22: ...ed in such an environment Emissions Test Compliance Electromagnetic Environment Guidance RF emissions CISPR 11 Group 1 The NewLife uses RF energy only for its internal function There fore its RF emiss...

Page 23: ...l on AC lines 2kV common mode on outdoor I O lines N A 1kV differential on AC lines N A Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interrupti...

Page 24: ...lectromag netic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz a...

Page 25: ...distance between the transmitting antenna and the ME EQUIPMENT OR ME SYSTEM may be reduced to 1m The 1 m test distance is permitted by IEC 61000 4 3 a For some services only uplink frequencies are in...

Page 26: ...the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80 MHz and 800 MHz the separation distance for the higher frequency range applies NOTE 2 These guidelines may not apply i...

Page 27: ...PN MN239 F User Manual ENG 27 NewLife Intensity 10 Notes...

Page 28: ...IRE Inc Please visit our website below for a full listing of trademarks Trademarks www caireinc com corporate trademarks Copyright 2022 CAIRE Inc All Rights Reserved CAIRE Inc reserves the right to di...

Reviews: