background image

1

20

Lib
er

at

ors

XI

Troubleshooting Charts

Symptom

Probable Cause

Corrective Action

1)

Unable  to  start  fill   

or    excessively                                             

l ong     fill   times

a)

Transfer line not engaged fully on 

Reservoir 

 

QDV

Make sure the QDV on the transfer line and reservoir are properly aligned and ensure 

that 

    a downward

 force is being applied to the transfer line assembly.

b)

Low source pressure

Verify 

   

  

that 

  

   

pressure

   

 

  

from 

   

the

    

 source

     

 is

    

 

  

within

    

 

  

the

    

 

  

35-50 psi

      

 range 

  

to 

  

fill 

  

the 

  

reservoir.

c)

Vent 

     

valve 

    

not 

     

open 

    

or

    

 is 

     

blocked

Ensure 

  

that 

  

the 

  

vent

  

 valve 

  

is 

  

able 

  

to 

  

be

  

 turned 

  

to 

  

the 

  

fully

  

 open 

  

position.

d)

Allow LOX to saturate to proper pressure.

e)

Fill connector 

  

not 

  

opening 

properly

Check fill connector and cartridge 

  

assembly for damage; 

  

make sure 

    

fill 

connectors

  

  

fully 

  

engage.

f) Vent valve is obstructed

Inspect 

  

the 

  

valve 

  

for 

  

blockages 

  

and 

  

verify 

  

that 

  

flow 

  

passes 

  

through 

  

during 

  

  

fill.  

Clean 

  

by

    

 blowing out with compressed gas or replace parts if necessary.

g)

Leak in the system

Check the reservoir for leaks (RP16) and repair if needed.

2)

Liquid  leaks  from 

the  coupled  QDVs 

during the fill

a)

Worn or damaged lip seal

Replace the QDV lip seal (RP19)

3)

Unable 

    

to 

disconnect 

     

the 

transfer 

  

line 

from 

  

the 

reservoir 

  

after 

 

   

fill

a)

Pop-off assembly not being utilized 

(Does not apply to Liberator 10)

Ensure 

 

that 

 

the 

 

pop-off

 

 assembly 

 

on

 

 t

 

he 

 

reservoir 

  

is 

  

being 

  

used.  

  

Do 

  

not 

    

use 

force

  

 to 

  

separate 

 

the QDVs.

b)

QDVs are frozen together

Leave the units coupled with the vent valve closed and let them sit until  they 

warm up enough to disconnect.   Always ensure that male and female  QDV’s  are 

cleaned

      

 and 

         

dried 

      

prior

      

 to each fill.

4)

Liquid  leaks  from 

the  QDV  poppet 

after 

        

f

 

illing

a)

Ice 

   

  

crystal 

  

preventing 

   

the 

   

QDV 

from closing properly.

Engage   and   disengage   the   transfer   line   onto   the   reservoir   several   times   to   dislodge the 

ice   crystal .  Always    be    sure    that   the   male   and   female      QDVs   are   wiped   clean   and   dry    

before  filling.

b)

Dirty

 

 or

 

 damaged 

 

QDV 

 

p

 

oppet

Replace the QDV cartridge (RP20) or the entire QDV assembly (RP21)

5)

Excessive  venting 

from  relief  valves 

(hissing

 

 

   

sound)

a)

Saturation pressure too high.

Inspect the saturation pressure of the reservoir used for filling.  Allow at

  

 least 

  

30 

    

minutes 

  

at no flow for the portable to saturate properly. 

b)

Relief valve frozen open

Allow the portable to warm and thaw.  Attempt to re-fill the portable.

c)

Faulty relief valve

Test the relief valve (RP13) and replace if necessary (RP14)

d)

Partial            or                   complete               loss                    of                       vacuum

Conduct    the    NER    test    (RP32)    and    return    the    unit    to    CAIRE,    Inc.    if    necessary.

6)

No

  

 flow 

 

at 

oxygen 

 

outlet

a)

Reservoir is empty

Check  the  contents  indicator/level  gauge  and  fill  the  reservoir  if  needed.

b)

Flow control valve turned off

Ensure the flow control knob is not in the off (“0”) position.

c)

Nasal 

            

cannula 

               

kinked 

                

or

               

 

  

disconnected

Ensure proper nasal cannula functionality and positioning

f)

Leak in the system

Perform a leak check on the plumbing (RP16).  Repair leaks as necessary.

g)

Relief valve is open

Ensure 

  

that

  

 there 

  

is

  

 no 

  

venting 

  

from

  

 the 

  

relief

  

 valves.  If 

  

there 

  

is 

  

refer 

  

to 

  

the 

corrective 

  

actions 

  

for 

 

“Excessive venting from relief valves (hissing sound)”

.

h)

Vent valve is open

Ensure that the

 

vent valve is fully closed.

i)

FCV inlet filter is obstructed

Clean or replace (RP30) the filter screen.

j)

Blockage     i n     the     liquid  withdrawal circuit

Check 

   

the

  

 warming 

  

coils 

  

and 

  

withdrawal

   

 tubes 

   

for 

   

blockages.  

   

Replace

   

 if 

   

necessary.

k)

FCV Faulty

Replace the FCV (RP30)

7)

Low  flow  at  oxygen 

outlet  on  all  LPM 

settings

a)

Nasal 

    

cannula 

    

kinked 

   

or 

    

leaking

Inspect the functionality of the nasal cannula.

b)

Saturation pressure is too low

Inspect

   

 the

    

 saturation

   

 pressure 

   

of 

   

the 

   

reservoir.  

   

Allow 

   

at

   

 least 

   

30 

   

minutes 

   

at 

no 

   

flow

   

 for

   

 the

   

 reservoir to saturate properly. 

c)

Leak in the system

Perform 

 

a

 

 leak 

 

check 

 

on the plumbing (RP16).  Repair 

 

leaks 

 

as 

 

necessary.

Source 

  

tank 

  

is 

  

either 

  

under 

or 

over-saturated

Liberators 

Technical Service Manual • PN 13350704 Rev 

K

 

7

Summary of Contents for Liberator G4 Series

Page 1: ...Liberators G4 Technical Service Manual...

Page 2: ...ic Rim Phone 770 721 7759 Fax 770 721 7758 Europe Phone 44 0 1189 367060 Fax 44 118 9799245 For Customer or Technical Service needs product assistance ordering part numbers specifications unexpected e...

Page 3: ...ption of a condition that can result in equipment or component damage NOTE A statement containing information important enough to emphasize or repeat ITEM Item numbers used throughout this manual are...

Page 4: ...ry of Operation 11 VII Unpacking and Setup 14 VIII Operation 15 IX Maintenance 17 X Troubleshooting 19 XI Troubleshooting Charts 20 23 XII Repair Procedures 24 XIII Service Tools Equipment Supplies 38...

Page 5: ...tempera ture and regulates the flow of oxygen gas to the patient Any water that condenses on the cold coils is routed into the con densate bottle An economizer regulator is utilized to con serve LOX...

Page 6: ...bar 23 2 psi 1 6 bar 23 2 psi Primary Relief Valve Reseat Pressure 1 4 bar 20 5 psi 1 4 bar 20 5 psi 1 4 bar 20 5 psi 1 4 bar 20 5 psi 1 4 bar 20 5 psi Secondary Relief Valve Opening Pressure 2 1 bar...

Page 7: ...as Association as an oxygen concentration above 23 In an oxygen enriched atmosphere flammable items may burn vigorously and may ex plode Certain items considered non combustible in air may burn rapidl...

Page 8: ...s are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11 Class B Harmonic emissions The Liberator is suitable for use in all establishments incl...

Page 9: ...input output lines No data input output lines 1 kV line s Surge to line s Not Applicable Not Applicable IEC 61000 4 5 2 kV line s DC powered device to earth Voltage dips 5 UT 95 dip short interruptio...

Page 10: ...nce level in each frequency rangeb Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These gui...

Page 11: ...2 3 P 0 01 0 12 m 0 12 m 0 23 m 0 1 0 38 m 0 38 m 0 73 m 1 1 2 m 1 2 m 2 3 m 10 3 8 m 3 8 m 7 3 m 100 12 m 12 m 23 m For transmitters rated at a maximum output power not listed above the recommended...

Page 12: ...e is substantially higher than the Liberator operating pressure oversaturated 2 Saturation pressure of the liquid oxygen in the fill source is substantially lower than the Liberator operating pressure...

Page 13: ...the bypass tee to the breathing coil the flow control valve and then the patient As the pressure in the container increases over 1 4 bar 20 psig the economizer valve opens and the cycle repeats mainta...

Page 14: ...e meter is powered by an internal battery offering battery life of 5 years or more at 30 cycles per day The meter has a low battery LOW BATT indicator which signals the need for bat tery replacement T...

Page 15: ...vel meter by depressing the operate button LEDs will light displaying the level of liquid oxygen If the unit is empty only the leftmost LED should light If it does not or if other erroneous indication...

Page 16: ...transfer line adapter to transfer line d Fully open liquid valve on fill source e Purge transfer line for a minimum of 5 seconds ensuring gas is safely piped away from operator i Connect transfer hose...

Page 17: ...tongs or tweezers 2 Wipe off the exterior of the unit with the cleaning solution us ing a lint free cloth The cloth should be damp but not drip ping Be as thorough as possible 3 Discard the cleaning c...

Page 18: ...nual NOTE See the Troubleshooting and Repair section of the cor responding technical manual on the detailed procedures for the tests mentioned below 1 Visual Inspection a Remove any LOX prior to maint...

Page 19: ...cator functionality d Presence of all required labels e Cryogenic reservoir damage dents dings f If LOX is still present in the unit inspect for heavy frost or condensation on the exterior of the unit...

Page 20: ...mbly RR 31 RP18 Vent Valve RR 31 RP19 Twist Lock QDV Lip Seal RR 31 RP20 Twist Lock QDV RR 32 RP21 Push on QDV RR 32 RP22 Push on Portable Pop Off Assembly 32 RP23 Empty Unit 33 RP24 Warm Unit 33 RP25...

Page 21: ...e QDVs are wiped clean and dry before filling b Dirty or damaged QDV poppet Replace the QDV cartridge RP20 or the entire QDV assembly RP21 5 Excessive venting from relief valves hissing sound a Satura...

Page 22: ...ionsfor Excessive venting from relief valves hissing sound f Partial or complete loss of vacuum ConducttheNERtest RP32 andreturntheunittoCAIRE Inc ifnecessary 10 Unit will not maintain acceptable syst...

Page 23: ...22 Liberators XI Troubleshooting Charts 8 Liberators Technical Service Manual PN 13350704 Rev K...

Page 24: ...e Troubleshooting Chart Start at the upper left corner The top line shows the steps of routine maintenance Unless otherwise noted by the arrows the flow through the chart is down or to the right Liber...

Page 25: ...attery WARNING Make sure the unit is empty and vent valve is open before replacing any component except shroud assembly compo nents or Lip Seals NOTE When assembling new compression fittings tighten 1...

Page 26: ...n more accurate but re quires the dewar be vent full with LOX To start either procedure calibration mode must first be entered There is no need to per form both procedures for any given calibration Pr...

Page 27: ...el 20 30 37 45 60 LED Min Max Min Max Min Max Min Max Min Max 1 17 6 20 6 22 0 24 7 22 7 27 9 24 9 31 3 34 2 38 1 2 20 6 23 5 24 7 28 9 27 9 33 1 31 3 37 6 38 1 46 3 3 23 5 26 5 28 9 33 2 33 1 38 3 37...

Page 28: ...7 Range The new full cap value is less than empty cap value When calibrating full Empty unit and recalibrate TABLE 6 Calibration Errors TABLE 7 Operation Errors LEDs Name Reason Notes Solution 2 4 6 H...

Page 29: ...e three bolts Item 31 that run through the FCV bracket Item 39 and through the manifold plate into the top of the dewar g Remove two remaining manifold mounting bolts Item 31 not visible in Figure 12...

Page 30: ...l holding signal wire against the ca pacitance probe c Unsolder feed thru wire Item A from probe Item B d Loosen feed thru nut Item C and remove harness e To replace manifold harness reverse above pro...

Page 31: ...Item 28 should open audible venting and or bubbling of leak test solution i Decrease pressure to 1 3 bar 19 psig Test relief valve with leak test solution A minimal amount of leakage bubbling is acce...

Page 32: ...o humidifier adapter Item 26 on FCV outlet Open FCV Item 24 to 6 LPM setting d Assemble oxygen regulator pneumatic hose Item C with DISS fittings and male pneumatic test adapter Item D Connect as semb...

Page 33: ...em 88 on tool f Install retaining ring Item 87 in QDV body Item15 by turning tool counter clockwise while pushing in on tool Torque retaining ring Item 87 to 225 280 N cm 20 25 in lbs RP17 Warming and...

Page 34: ...e static seal Item 67 if needed FIGURE 21 Top Fill Push on QDV Assembly WARNING Liberator must be empty and vented before starting procedure WARNING Liberator must be empty and vented before starting...

Page 35: ...g 4 Set Liberator FCV Item 24 to 6 LPM setting 5 Continue until Liberator is empty RP24 Warm Unit Figure 23 a Liberator may be allowed to sit FCV off vent closed a mini mum of 48 hours after emptying...

Page 36: ...alues within tolerances listed in Table 9 or unit fails flow rate test NOTE Flow rates will only be accurate if the reservoir is filled properly Reference the filling section of this manual for the co...

Page 37: ...ected to normal temperature and pressure 70 F 14 7 psig or 21 C and 1 01 Bar The above acceptable flow rates are intended for field testing only of flow control valves to account for variation in elev...

Page 38: ...ator Item 35 c Follow RP29 to ensure proper settings on the replacement regulator d Reverse remainder of procedure to install new properly ad justed regulator RP29 Economizer Regulator Test a Follow s...

Page 39: ...en in a full Liberator 45 reservoir 1 Empty weight of LIB 45 Reservoir 24 95 kg 55 00 lbs 2 Filled unit with properly saturated liquid oxygen 3 Full weight of LIB 45 Reservoir 74 99 kg 165 32 lbs 4 74...

Page 40: ...uper Flex Liquid Oxygen Transfer Line 6 97406630 Dual Fill Head Tee 13350704 Service Manual 10661515 Conversion Kit TF to SF 10660361 Conversion Kit SF to Dual 10660344 Conversion Kit TF to Dual 10661...

Page 41: ...40 Liberators Parts Price List XIV Contact Customer Service or visit www cairemedical com to obtain your parts list Liberators Technical Service Manual PN 13350704 Rev K...

Page 42: ...Fax 44 0 1344 429224 4 Mail or fax the completed purchase order for confirmation to CAIRE Inc 2200 Airport Industrial Dr Ste 500 Ball Ground GA 30115 USA CAIRE Medical Ltd Unit 6 Ashville Way Wokingha...

Page 43: ...ntenance performed Restocking Policy If it becomes necessary to cancel an order with CAIRE Inc af ter the shipment has been received use the following Restock Policy procedure 1 Notify the Customer Se...

Page 44: ...Wokingham Berkshire RG41 2PL United Kingdom Ph 44 0 1189 367060 Fax 44 0 0118 979 9245 Copyright CAIRE Inc 2019 CAIRE Inc reserves the right to discontinue its products or change the prices materials...

Reviews: