background image

42

Lib
er

at

ors

Return & Restocking Policy

XVI

When a Liberator is received, it should be inspected immediately, 

as outlined in Section VII, Unpacking and Setup Instructions.
If a problem with the unit should be encountered, reference 

should be made to the Troubleshooting Chart in Section X, page 

12-13. If these procedures do not provide a solution for the prob-

lem, the following steps should be taken:

1.

Call CAIRE Inc. Customer Service. State the problem with

the unit. If it is determined that the problem cannot be

solved by the distributor, a Return Material Authorization

(RMA) number will be assigned to the unit or part(s). If a

Purchase Order Number is to be referenced, please give this

number to the Customer Service Representative at that time.

2.

Carefully package the parts, or repack the unit in its original

shipping container, precisely as shipped.

3.

Write the Return Authorization Number on the top of the

shipping container.

4.

Return the unit or parts by professional carrier to:

CAIRE Inc.
2205 Airport Drive
Ball Ground, GA 30107

C

AIRE

 Medical

 Ltd.

Unit 6, Ashville Way Wokingham
Berkshire, RG41 2PL United Kingdom
Ph +44 (0) 1189 367060 Fax +44(0) 0118 979 9245

All equipment returned to CAIRE Inc. must be shipped  

     “prepaid”.

When the defective item(s) is received at CAIRE Inc.,  it will be 

serviced and returned to the distributor as soon as possible. A 

copy of the “Repair Cost Sheet” will be enclosed giving a detailed 

listing of any maintenance performed.

Restocking Policy

If it becomes necessary to cancel an order with CAIRE Inc. af-

ter the shipment has been received, use the following “Restock 

Policy” procedure:

1. Notify the Customer Service Department at CAIRE Inc. us-

ing the toll-free number. When contacting Customer Service

personnel, it will be necessary to relay the following infor-

mation:

a. State the quantity and description of equipment to be re-

turned.

b. Give the Serial Number of each unit to be returned.

c. State the equipment purchase date.

2.

A RMA number will be issued in the name of the distributor

 

by CAIRE, Inc. for the equipment to be returned. When the

 

equipment is shipped to the factory, the RMA number must

 

appear on the packing slip.

3.

All equipment must be returned “prepaid” to:

CAIRE 

Inc

.

2205 Airport Drive
Ball Ground, GA 30107

C

AIRE

 Medical

, Ltd.

Unit 6, Ashville Way Wokingham
Berkshire, RG41 2PL United Kingdom
Ph +44 (0) 1189 367060 Fax +44(0) 0118 979 9245

4.

Finally, a “Credit Memo”, minus a 15% restocking fee, will

 

be

 

issued to the distributor when all equipment has been re-

ceived, inspected, and restocked by CAIRE, Inc.

Return of Unused Non-Defective Merchandise

CAIRE Inc.,  at its discretion, charges a 15% restocking fee for 

unused non-defective merchandise that is returned. An RMA 

number must be obtained from CAIRE Inc. Customer Service 

prior to return of any goods. Merchandise cannot be returned for 

credit after sixty (60) days. Customer to pay all freight charges. 

Tracking capability and insurance on all returned goods is ad-

vised. CAIRE Inc. will not be responsible for misdirected ship-

ments.

Liberators Technical Service Manual • PN 13350704 Rev 

K

 

Summary of Contents for Liberator G4 Series

Page 1: ...Liberators G4 Technical Service Manual...

Page 2: ...ic Rim Phone 770 721 7759 Fax 770 721 7758 Europe Phone 44 0 1189 367060 Fax 44 118 9799245 For Customer or Technical Service needs product assistance ordering part numbers specifications unexpected e...

Page 3: ...ption of a condition that can result in equipment or component damage NOTE A statement containing information important enough to emphasize or repeat ITEM Item numbers used throughout this manual are...

Page 4: ...ry of Operation 11 VII Unpacking and Setup 14 VIII Operation 15 IX Maintenance 17 X Troubleshooting 19 XI Troubleshooting Charts 20 23 XII Repair Procedures 24 XIII Service Tools Equipment Supplies 38...

Page 5: ...tempera ture and regulates the flow of oxygen gas to the patient Any water that condenses on the cold coils is routed into the con densate bottle An economizer regulator is utilized to con serve LOX...

Page 6: ...bar 23 2 psi 1 6 bar 23 2 psi Primary Relief Valve Reseat Pressure 1 4 bar 20 5 psi 1 4 bar 20 5 psi 1 4 bar 20 5 psi 1 4 bar 20 5 psi 1 4 bar 20 5 psi Secondary Relief Valve Opening Pressure 2 1 bar...

Page 7: ...as Association as an oxygen concentration above 23 In an oxygen enriched atmosphere flammable items may burn vigorously and may ex plode Certain items considered non combustible in air may burn rapidl...

Page 8: ...s are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11 Class B Harmonic emissions The Liberator is suitable for use in all establishments incl...

Page 9: ...input output lines No data input output lines 1 kV line s Surge to line s Not Applicable Not Applicable IEC 61000 4 5 2 kV line s DC powered device to earth Voltage dips 5 UT 95 dip short interruptio...

Page 10: ...nce level in each frequency rangeb Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These gui...

Page 11: ...2 3 P 0 01 0 12 m 0 12 m 0 23 m 0 1 0 38 m 0 38 m 0 73 m 1 1 2 m 1 2 m 2 3 m 10 3 8 m 3 8 m 7 3 m 100 12 m 12 m 23 m For transmitters rated at a maximum output power not listed above the recommended...

Page 12: ...e is substantially higher than the Liberator operating pressure oversaturated 2 Saturation pressure of the liquid oxygen in the fill source is substantially lower than the Liberator operating pressure...

Page 13: ...the bypass tee to the breathing coil the flow control valve and then the patient As the pressure in the container increases over 1 4 bar 20 psig the economizer valve opens and the cycle repeats mainta...

Page 14: ...e meter is powered by an internal battery offering battery life of 5 years or more at 30 cycles per day The meter has a low battery LOW BATT indicator which signals the need for bat tery replacement T...

Page 15: ...vel meter by depressing the operate button LEDs will light displaying the level of liquid oxygen If the unit is empty only the leftmost LED should light If it does not or if other erroneous indication...

Page 16: ...transfer line adapter to transfer line d Fully open liquid valve on fill source e Purge transfer line for a minimum of 5 seconds ensuring gas is safely piped away from operator i Connect transfer hose...

Page 17: ...tongs or tweezers 2 Wipe off the exterior of the unit with the cleaning solution us ing a lint free cloth The cloth should be damp but not drip ping Be as thorough as possible 3 Discard the cleaning c...

Page 18: ...nual NOTE See the Troubleshooting and Repair section of the cor responding technical manual on the detailed procedures for the tests mentioned below 1 Visual Inspection a Remove any LOX prior to maint...

Page 19: ...cator functionality d Presence of all required labels e Cryogenic reservoir damage dents dings f If LOX is still present in the unit inspect for heavy frost or condensation on the exterior of the unit...

Page 20: ...mbly RR 31 RP18 Vent Valve RR 31 RP19 Twist Lock QDV Lip Seal RR 31 RP20 Twist Lock QDV RR 32 RP21 Push on QDV RR 32 RP22 Push on Portable Pop Off Assembly 32 RP23 Empty Unit 33 RP24 Warm Unit 33 RP25...

Page 21: ...e QDVs are wiped clean and dry before filling b Dirty or damaged QDV poppet Replace the QDV cartridge RP20 or the entire QDV assembly RP21 5 Excessive venting from relief valves hissing sound a Satura...

Page 22: ...ionsfor Excessive venting from relief valves hissing sound f Partial or complete loss of vacuum ConducttheNERtest RP32 andreturntheunittoCAIRE Inc ifnecessary 10 Unit will not maintain acceptable syst...

Page 23: ...22 Liberators XI Troubleshooting Charts 8 Liberators Technical Service Manual PN 13350704 Rev K...

Page 24: ...e Troubleshooting Chart Start at the upper left corner The top line shows the steps of routine maintenance Unless otherwise noted by the arrows the flow through the chart is down or to the right Liber...

Page 25: ...attery WARNING Make sure the unit is empty and vent valve is open before replacing any component except shroud assembly compo nents or Lip Seals NOTE When assembling new compression fittings tighten 1...

Page 26: ...n more accurate but re quires the dewar be vent full with LOX To start either procedure calibration mode must first be entered There is no need to per form both procedures for any given calibration Pr...

Page 27: ...el 20 30 37 45 60 LED Min Max Min Max Min Max Min Max Min Max 1 17 6 20 6 22 0 24 7 22 7 27 9 24 9 31 3 34 2 38 1 2 20 6 23 5 24 7 28 9 27 9 33 1 31 3 37 6 38 1 46 3 3 23 5 26 5 28 9 33 2 33 1 38 3 37...

Page 28: ...7 Range The new full cap value is less than empty cap value When calibrating full Empty unit and recalibrate TABLE 6 Calibration Errors TABLE 7 Operation Errors LEDs Name Reason Notes Solution 2 4 6 H...

Page 29: ...e three bolts Item 31 that run through the FCV bracket Item 39 and through the manifold plate into the top of the dewar g Remove two remaining manifold mounting bolts Item 31 not visible in Figure 12...

Page 30: ...l holding signal wire against the ca pacitance probe c Unsolder feed thru wire Item A from probe Item B d Loosen feed thru nut Item C and remove harness e To replace manifold harness reverse above pro...

Page 31: ...Item 28 should open audible venting and or bubbling of leak test solution i Decrease pressure to 1 3 bar 19 psig Test relief valve with leak test solution A minimal amount of leakage bubbling is acce...

Page 32: ...o humidifier adapter Item 26 on FCV outlet Open FCV Item 24 to 6 LPM setting d Assemble oxygen regulator pneumatic hose Item C with DISS fittings and male pneumatic test adapter Item D Connect as semb...

Page 33: ...em 88 on tool f Install retaining ring Item 87 in QDV body Item15 by turning tool counter clockwise while pushing in on tool Torque retaining ring Item 87 to 225 280 N cm 20 25 in lbs RP17 Warming and...

Page 34: ...e static seal Item 67 if needed FIGURE 21 Top Fill Push on QDV Assembly WARNING Liberator must be empty and vented before starting procedure WARNING Liberator must be empty and vented before starting...

Page 35: ...g 4 Set Liberator FCV Item 24 to 6 LPM setting 5 Continue until Liberator is empty RP24 Warm Unit Figure 23 a Liberator may be allowed to sit FCV off vent closed a mini mum of 48 hours after emptying...

Page 36: ...alues within tolerances listed in Table 9 or unit fails flow rate test NOTE Flow rates will only be accurate if the reservoir is filled properly Reference the filling section of this manual for the co...

Page 37: ...ected to normal temperature and pressure 70 F 14 7 psig or 21 C and 1 01 Bar The above acceptable flow rates are intended for field testing only of flow control valves to account for variation in elev...

Page 38: ...ator Item 35 c Follow RP29 to ensure proper settings on the replacement regulator d Reverse remainder of procedure to install new properly ad justed regulator RP29 Economizer Regulator Test a Follow s...

Page 39: ...en in a full Liberator 45 reservoir 1 Empty weight of LIB 45 Reservoir 24 95 kg 55 00 lbs 2 Filled unit with properly saturated liquid oxygen 3 Full weight of LIB 45 Reservoir 74 99 kg 165 32 lbs 4 74...

Page 40: ...uper Flex Liquid Oxygen Transfer Line 6 97406630 Dual Fill Head Tee 13350704 Service Manual 10661515 Conversion Kit TF to SF 10660361 Conversion Kit SF to Dual 10660344 Conversion Kit TF to Dual 10661...

Page 41: ...40 Liberators Parts Price List XIV Contact Customer Service or visit www cairemedical com to obtain your parts list Liberators Technical Service Manual PN 13350704 Rev K...

Page 42: ...Fax 44 0 1344 429224 4 Mail or fax the completed purchase order for confirmation to CAIRE Inc 2200 Airport Industrial Dr Ste 500 Ball Ground GA 30115 USA CAIRE Medical Ltd Unit 6 Ashville Way Wokingha...

Page 43: ...ntenance performed Restocking Policy If it becomes necessary to cancel an order with CAIRE Inc af ter the shipment has been received use the following Restock Policy procedure 1 Notify the Customer Se...

Page 44: ...Wokingham Berkshire RG41 2PL United Kingdom Ph 44 0 1189 367060 Fax 44 0 0118 979 9245 Copyright CAIRE Inc 2019 CAIRE Inc reserves the right to discontinue its products or change the prices materials...

Reviews: