PN 21157307-C4 B | User Manual 11 - ENG
HELiOS
Safety
WARNING:
PORTABLE RF COMMUNICATIONS EQUIPMENT (INCLUDING PERIPHERALS
SUCH AS ANTENNA CABLES AND EXTERNAL ANTENNAS) SHOULD BE USED NO CLOSER
THAN 30 CM (12 INCHES) TO ANY PART OF THE HELIOS, INCLUDING CABLES SPECIFIED
BY THE MANUFACTURER. OTHERWISE, DEGRADATION OF THE PERFORMANCE OF THIS
EQUIPMENT COULD RESULT.
WARNING: USE OF ACCESSORIES, TRANSDUCERS AND CABLES OTHER THAN THOSE SPECIFIED OR
PROVIDED BY THE MANUFACTURER OF THIS EQUIPMENT COULD RESULT IN INCREASED ELECTROMAG-
NETIC EMISSIONS OR DECREASED ELECTROMAGNETIC IMMUNITY OF THIS EQUIPMENT AND RESULT IN
IMPROPER OPERATION.
WARNING: USE OF THIS EQUIPMENT ADJACENT TO OR STACKED WITH OTHER EQUIPMENT SHOULD
BE AVOIDED BECAUSE IT COULD RESULT IN IMPROPER OPERATION. IF SUCH USE IS NECESSARY, THIS
EQUIPMENT AND THE OTHER EQUIPMENT SHOULD BE OBSERVED TO VERIFY THAT THEY ARE OPERAT-
ING NORMALLY.
Caution: Medical Electrical Equipment needs special precautions regarding Electromagnetic
compatibility (EMC) and needs to be installed and put into service according to the EMC infor-
mation provided in this manual.
Caution: Portable and mobile radio frequency (RF) communications equipment can affect
Medical Electrical Equipment.
Table 1
Guidance and Manufacturer’s declaration—electromagnetic emissions
The HELiOS Reservoir is intended for use in the electromagnetic environment specified below. The
customer or the user of the HELiOS Reservoir should assure that it is used in such an environment.
Emissions test
Compliance Electromagnetic environment—guidance
RF emissions
Group 1
The HELiOS uses RF energy only for internal function.
CISPR 11
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emissions CISPR 11
Class B
Harmonic emissions
The HELiOS is suitable for use in all establishments, including
IEC 61000-3-2
Not applicable domestic establishments and those directly connected to the
Voltage fluctuations/
public low-voltage power supply network that supplies
flicker emissions
Not applicable buildings used for domestic purposes.
IEC 61000-3-3