Companion Portable Technical Service Manual • PN 20562220 Rev B
6
Companion P
or
tables
III
Safety
Table 2
Guidance and Manufacturer’s declaration—electromagnetic emissions
The Companion 1000/T is intended for use in the electromagnetic environment specified below. The customer or the user of the
Companion 1000/T should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment—guidance
RF emissions
Group 1
The Companion 1000/T uses RF energy only for internal function.
CISPR 11
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emissions CISPR 11
Class B
Harmonic emissions
The Companion 1000/T is suitable for use in all establishments,
IEC 61000-3-2
Not applicable
including domestic establishments and those directly connected to the
Voltage fluctuations/
public low-voltage power supply network that supplies buildings used
flicker emissions
Not applicable
for domestic purposes.
IEC 61000-3-3
• Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into service according
to the EMC information provided in this manual.
• Portable and mobile RF communications equipment can affect Medical Electrical Equipment.
• The use of Accessories, transducers, and cables other than those specified, with the exception of transducers and cables sold by
the Manufacturer of this device as replacement parts for internal components, may result in increased Emissions or decreased
Immunity of the Companion 1000/T.
• The Companion 1000/T should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is
necessary, the Companion 1000/T should be observed to verify normal operation in the configuration in which it will be used.